Fluke 700G31 Pressure Gauge Calibrator 10000psi 4097622

£1,031.34 £859.45

Fluke 700G Series Technical Specifications

Model Range Resolution Accuracy Burst Pressure
Fluke 700G01 -10 to +10 inH2O
-20 mbar to 20 mbar
0.001 inH2O,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
3 psi,
210 mbar
Fluke 700G02 -1 to +1 psi,
-70 mbar to 70 mbar
0.0001 psi,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
5 psi,
350 mbar
Fluke 700G04 -14 to 15 psi,
-0.97 to 1 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
60 psi,
4 bar
Fluke 700G05 -14 to 30 psi,
-0.97 to 2 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
120 psi,
8 bar
Fluke 700G06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
400 psi,
26 bar
Fluke 700G27 -12 to 300 psi,
-0.83 to 20 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
1200 psi,
80 bar
Fluke 700G07 -12 to 500 psi,
-0.83 to 34 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
2000 psi,
138 bar
Fluke 700G08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
4000 psi,
266 bar
Fluke 700G10 -14 to 2000 psi,
-.97 bar to 140 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
8000 psi,
550 bar
Fluke 700G29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke 700G30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke 700G31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
20000 psi,
1035 bar
Fluke 700GA4 0 to 15 psia,
0 to 1 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 60 psi,
4 bar
Fluke 700GA5 0 to 30 psia
0 to 2 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 120 psi,
8 bar
Fluke 700GA6 0 to 100 psia,
0 to 6.9 bar absolute
0.01 psi,
0.0001 bar
±0.05% of range 400 psi,
27 bar
Fluke 700GA27 0 to 300 psia,
0 to 20 bar absolute
0.01 psi,
0.001 bar
±0.05% of range 1200 psi,
80 bar
Fluke 700RG05 -14 to +30 psi,
-0.97 to +2 bar
0.001 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.05% of range
90 psi,
6 bar
Fluke 700RG06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
400 psi,
27 bar
Fluke 700RG07 -12 to 500 psi,
-0.83 to 34
0.01 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
2000 psi,
138 bar
Fluke 700RG08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
4000 psi,
275 bar
Fluke 700RG29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
10000 psi,
690 bar
Fluke 700RG30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
15000 psi,
1035 bar
Fluke 700RG31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
20000 psi,
1380 bar
Media Compatability        
Gas Types 10 inH2O, 1, 15, 30 psi: any clean dry non-corrosive gas
100, 300, 500, 1000 psi: any liquids or gases compatible with 316 stainless steel and above 1000 psi: any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel.
Mechanical Specifications        
Dimensions (H x W x D) 12.7 x 11.4 x 3.7cm (5 x 4.5 x 1.5in)
Pressure Connection ¼ in NPT male
Housing/Weight Cast ZNAL 0.56kg (1.22lbs) with holster
Display 5 ½ digits, 16.53mm (0.65in) high 20-segment bar graph, 0 to 100%
Power 3 x AA alkaline batteries
Battery Life 1500 hours without backlight (continuous on), 2000 hours at slow sample rate

Description

Fluke 700G31* Pressure Gauge Calibrator 10000psi 4097622

-14 to 10000 psi
-0.97 to 690 bar

Portable, high-quality round pressure gauge calibrator for fast and accurate calibration test results 

Fluke 700G Series precision pressure gauge calibrators are portable, high quality calibrators offering fast and accurate calibration test results. The 700G models are easy-to-use and of rugged construction for reliable performance in different environments. They are compatible with most hydraulic and pneumatic test pumps.

*Important Calibration Information: 

This instrument comes complete with the manufacturer’s calibration certificate. Please be aware that these certificates are generic and may also be time lapsed.[Text Wrapping Break]If you would prefer a personalised full traceable calibration certificate applied to the equipment before dispatch, then please select this option from the drop-down menu above, an additional calibration charge will apply. 

Included:

  • Manual on CD-ROM in 14 languages
  • Traceable certificate of calibration
  • 1/4˝ NPT to 1/4˝ ISO adapter

Key Features:

Designed for field pressure calibration, the Fluke 700G Series Pressure Gauge Calibrators give you the performance and durability that you have come to expect in Fluke pressure calibration products while being simple and easy-to-use. Whether you’re looking to connect to either a hydraulic or pneumatic test pump this series are designed to be a complete calibration solution with:

  • The 700 series includes 23 pressure gauges from ±10 inH2O/20 mbar to 10,000 psi/690 bar

With a standard accuracy of 0.05% of full scale and reference class gauges with accuracies of 0.04% you’re sure to find a gauge that’s the perfect fit for your application.

Pressure gauge data logging

The Fluke 700G series gauges and pump kits are designed for the pressure testing professional working on process pressure instrumentation. The pressure ranges of the 700G series gauges cover the ranges needed for process pressure calibration needs including Gage pressures +/- 10 inH20 (+/- 25 mbar) to 10,000 psi (0-690 bar), absolute pressures from 0 to 15 psia (0-1 bar) to 0-300 psi (0-20 bar) and reference class gauges from 0 to 30 PSI (0-2 bar) to 10,000 psi (0-690 bar).

  • Standard accuracy to 0.05% of full scale
  • Reference class gauges deliver 0.04% of reading measurement accuracy
  • Pressure gauge data logging for more than 8,493 pressure measurements to memory and uploads to a PC (with 700G/TRACK software)

Rugged design with intrinsically safe certification

  • Rugged sealed IP-67 enclosure with included protective holster
  • Allows you to work safely in classified areas known to include explosive gasses:
    • CSA: Class 1, Div2, Groups A-D rating
    • ATEX: II 3 G Ex ic IIB T6 Gc
  • Standard 3-year warranty

Easy-to-use pressure solution

Pressure gauge calibrators are the easiest to use of all pressure calibration tools. Power them up, connect to the pressure device you want to measure and see your measurement, it’s that easy. When used with a Fluke-700PTPK pneumatic or 700HTPK hydraulic pump kit the 700G becomes a complete pressure sourcing and measurement solution. The 700PTPK2 and 700HTPK2 premium pump kits offer easy connectivity to the gauge and device under test providing “no tools required” connections.

  • Complete pressure calibration solution when gauges are paired with special pump kits
    • Calibrate to 600 psi (20 bar) with the 700PTPK or 700PTPK2 pneumatic pump kit.
    • Calibrate to 10,000 psi (690 bar) with the 700HTPK or 700HTPK2 hydraulic pump kit.

Fluke 700G Pressure Gauge Calibrator Selector

700R Ranges: Temperature Compensation 0 e °C to 50 °C (32 °F to 122 °F) to rated accuracy.
For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add .005% FS/ °C
Model Range Resolution Accuracy Burst Pressure
Fluke-700G01 -10 to +10 inH2O
-20 mbar to 20 mbar
0.001 inH2O,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
3 psi,
210 mbar
Fluke-700G02 -1 to +1 psi,
-70 mbar to 70 mbar
0.0001 psi,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
5 psi,
350 mbar
Fluke-700G04 -14 to 15 psi,
-0.97 to 1 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
60 psi,
4 bar
Fluke-700G05 -14 to 30 psi,
-0.97 to 2 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
120 psi,
8 bar
Fluke-700G06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
400 psi,
26 bar
Fluke-700G27 -12 to 300 psi,
-0.83 to 20 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
1200 psi,
80 bar
Fluke-700G07 -12 to 500 psi,
-0.83 to 34 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
2000 psi,
138 bar
Fluke-700G08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
4000 psi,
266 bar
Fluke-700G10 -14 to 2000 psi,
-.97 bar to 140 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
8000 psi,
550 bar
Fluke-700G29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke-700G30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke-700G31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
20000 psi,
1035 bar
Fluke-700GA4 0 to 15 psia,
0 to 1 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 60 psi,
4 bar
Fluke-700GA5 0 to 30 psia
0 to 2 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 120 psi,
8 bar
Fluke-700GA6 0 to 100 psia,
0 to 6.9 bar absolute
0.01 psi,
0.0001 bar
±0.05% of range 400 psi,
27 bar
Fluke-700GA27 0 to 300 psia,
0 to 20 bar absolute
0.01 psi,
0.001 bar
±0.05% of range 1200 psi,
80 bar
Fluke-700RG05 -14 to +30 psi,
-0.97 to +2 bar
0.001 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.05% of range
90 psi,
6 bar
Fluke-700RG06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
400 psi,
27 bar
Fluke-700RG07 -12 to 500 psi,
-0.83 to 34
0.01 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
2000 psi,
138 bar
Fluke-700RG08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
4000 psi,
275 bar
Fluke-700RG29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
10000 psi,
690 bar
Fluke-700RG30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
15000 psi,
1035 bar
Fluke-700RG31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
20000 psi,
1380 bar
Media compatibility
  • 10 inH2O, 1, 15, 30 psi: any clean dry non-corrosive gas
  • 100, 300, 500, 1000 psi: any liquids or gases compatible with 316 stainless steel and above 1000 psi:
    any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel.
Mechanical Specifications
Dimensions (H x W x D) 12.7 x 11.4 x 3.7 cm (5 x 4.5 x 1.5 in)
Pressure connection ¼ in NPT Male
Housing/Weight Cast ZNAL 0.56 kg (1.22 lb) with holster
Display 5-1/2 digits, 16.53 mm (0.65 in) high 20-segment bar graph, 0 to 100%
Power Three (3) AA alkaline batteries
Battery life 1,500 hours without backlight (continuous on), 2,000 hours at slow sample rate


isswww and Fluke Instruments

Fluke Corporation is the world leader in the manufacture, distribution and service of electronic test tools and software.

Since its founding in 1948, Fluke has helped define and grow a unique technology market, providing testing and troubleshooting capabilities that have grown to mission-critical status in manufacturing and service industries. Every new manufacturing plant, office, hospital, or facility built today represents another potential customer for Fluke products.

From industrial electronic installation, maintenance and service, to precision measurement and quality control, Fluke tools help keep businesses and industries around the globe up and running. Typical customers and users include technicians, engineers, metrologists, medical-device manufacturers, and computer network professionals, people who stake their reputations on their tools, and use tools to help extend their personal power and abilities.

Fluke are a founding core supplier of ISSWWW and we enjoy a personal and professional relationship with them. Our sales staff and our technicians are Fluke trained and we hold perhaps the largest stock of their products in the country. Our customers know that when it comes to Fluke products, we have not only the best prices but the best advice and support services, reassuring when you need to make that all-important purchase.

 

Features & Specification

More Information
Key Features Absolute pressure measurement ranges 15 30 100 300 psia, Ranges: ±10 inH2O/20 mbar to 10000 psi/690 bar, Accuracy to 0.05 % of full scale, Easy to use rugged construction, CSA Class 1 Div 2 Groups A-D rating, ATEX rating: II 3 G Ex nA IIB T6, Log up to 8493 pressure measurements (requires software), ¼ in NPT Male
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Fluke Testers
Short Description

Fluke 700G Series Technical Specifications

Model Range Resolution Accuracy Burst Pressure
Fluke 700G01 -10 to +10 inH2O
-20 mbar to 20 mbar
0.001 inH2O,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
3 psi,
210 mbar
Fluke 700G02 -1 to +1 psi,
-70 mbar to 70 mbar
0.0001 psi,
0.001 mbar
Positive pressure ±0.1% of range,
Vacuum ±0.1% of range
5 psi,
350 mbar
Fluke 700G04 -14 to 15 psi,
-0.97 to 1 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
60 psi,
4 bar
Fluke 700G05 -14 to 30 psi,
-0.97 to 2 bar
0.001 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
120 psi,
8 bar
Fluke 700G06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
400 psi,
26 bar
Fluke 700G27 -12 to 300 psi,
-0.83 to 20 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
1200 psi,
80 bar
Fluke 700G07 -12 to 500 psi,
-0.83 to 34 bar
0.01 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
2000 psi,
138 bar
Fluke 700G08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
4000 psi,
266 bar
Fluke 700G10 -14 to 2000 psi,
-.97 bar to 140 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
8000 psi,
550 bar
Fluke 700G29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke 700G30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
10000 psi,
690 bar
Fluke 700G31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure ±0.05% FS
Vacuum ±0.1% FS
20000 psi,
1035 bar
Fluke 700GA4 0 to 15 psia,
0 to 1 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 60 psi,
4 bar
Fluke 700GA5 0 to 30 psia
0 to 2 bar absolute
0.001 psi,
0.0001 bar
±0.05% of range 120 psi,
8 bar
Fluke 700GA6 0 to 100 psia,
0 to 6.9 bar absolute
0.01 psi,
0.0001 bar
±0.05% of range 400 psi,
27 bar
Fluke 700GA27 0 to 300 psia,
0 to 20 bar absolute
0.01 psi,
0.001 bar
±0.05% of range 1200 psi,
80 bar
Fluke 700RG05 -14 to +30 psi,
-0.97 to +2 bar
0.001 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.05% of range
90 psi,
6 bar
Fluke 700RG06 -12 to 100 psi,
-0.83 to 6.9 bar
0.01 psi,
0.0001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
400 psi,
27 bar
Fluke 700RG07 -12 to 500 psi,
-0.83 to 34
0.01 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
2000 psi,
138 bar
Fluke 700RG08 -14 to 1000 psi,
-0.97 to 69 bar
0.1 psi,
0.001 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
4000 psi,
275 bar
Fluke 700RG29 -14 to 3000 psi,
-0.97 to 200 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
10000 psi,
690 bar
Fluke 700RG30 -14 to 5000 psi,
-0.97 to 340 bar
0.1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
15000 psi,
1035 bar
Fluke 700RG31 -14 to 10000 psi,
-0.97 to 690 bar
1 psi,
0.01 bar
Positive pressure,
±0.04% of reading +.01% of range,
Vacuum ±0.1% of range
20000 psi,
1380 bar
Media Compatability        
Gas Types 10 inH2O, 1, 15, 30 psi: any clean dry non-corrosive gas
100, 300, 500, 1000 psi: any liquids or gases compatible with 316 stainless steel and above 1000 psi: any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel.
Mechanical Specifications        
Dimensions (H x W x D) 12.7 x 11.4 x 3.7cm (5 x 4.5 x 1.5in)
Pressure Connection ¼ in NPT male
Housing/Weight Cast ZNAL 0.56kg (1.22lbs) with holster
Display 5 ½ digits, 16.53mm (0.65in) high 20-segment bar graph, 0 to 100%
Power 3 x AA alkaline batteries
Battery Life 1500 hours without backlight (continuous on), 2000 hours at slow sample rate

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA