Infrared Measurements
Infrared temperature range -30 °C to 900 °C (-22 °F to 1652 °F)
IR accuracy (calibration geometry with ambient temperature 23 °C ± 2 °C) ≥ 0°C: ± 1°C or ± 1 % of the reading, whichever is greater
≥-10 °C to <0 °C: ±2°C
<-10°C: ±3 °C

≥32°F: ± 2°F or ± 1% of the reading, whichever is greater
≥14 °F to <32 °F: ±4 °F
<14 °F: ±6 °F
IR repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater
Display resolution 0.1 °C / 0.1 °F
Distance: Spot 60:1 (calculated at 90 % energy)
Minimum spot size 19 mm
Laser sighting Offset dual laser, output <1 mW
Spectral response 8 μm to 14 μm
Response time (95 %) <500 ms
Emissivity Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials

 

Contact Measurements
K-type thermocouple input temperature range -270 °C to 1372 °C (-454 °F to 2501 °F)
K-type thermocouple input accuracy (with ambient temperature 23 °C ± 2 °C) <-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
K-type thermocouple 0.1 °C/0.1 °F
K-type thermocouple repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater

 

Measurement Options
Hi/Low alarms Audible and two-color visual
Min/Max/Avg/Dif Yes
Switchable celsius and fahrenheit Yes
Backlight Two levels, normal and extra bright for darker environments
Probe input K-type thermocouple simultaneous display of probe and IR temperature
Trigger lock Yes
Data storage 99 points
Display Dot matrix 98 x 96 pixels with function menus
Communication USB 2.0

 

K-Type Thermocouple Specifications
Measurement range (bead probe) -40 °C to 260 °C (-40 °F to 500 °F)
Accuracy ±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F)
Cable length 1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector and bead termination

 

General Specifications
Safety and compliance IEC 60825-1 
FDA Laser Class II 
IEC 61326-1 
CE Complaint 
CMC # 01120009
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Relative humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating altitude 2000 meters above mean sea level
Weight 0.322 kg (0.7099 lb)
Power 2 AA batteries
Battery life 8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on)

Description

Fluke 572-2 High Temperature Infrared Thermometer

With a rugged, easy-to-use, ergonomic design, the Fluke 575-2 can stand up to tough industrial, electrical, and mechanical environments

 

The Fluke 572-2 High-Temperature Infrared Thermometer is the one product you can use in high-temperature industrial environments all around the world. Whether you work in power utility, metal refining and smelting, glass, cement or petrochemical environments, the 572-2 allows you to carry the most trusted name in test tools anywhere you need accurate, high-temperature and high distance-to-spot measurements.


Fluke 572-2 Comes Supplied With:

  • K-type thermocouple bead probe
  • Durable hard case
  • USB 2.0 computer interface cable

With a straight-forward user interface and soft-key menus, the Fluke 572-2 makes even complex measurements easy. Quickly navigate and adjust emissivity, start data logging, or turn on and off alarms, with just a few pushes of a button.

 

Fluke 572-2 High-Temperature Handheld Infrared Thermometer

  • Measure -30 °C to 900 °C (-22 °F to 1652 °F)
  • 60:1 distance to spot ratio with dual laser sighting for fast, accurate targeting
  • Multi-language interface (user select)
  • Current Temperature plus MAX, MIN, DIF, AVG temperature displays
  • Compatible with standard mini-connector K-type thermocouples, including ones you already own and have installed
  • Adjustable emissivity and predefined emissivity table
  • Infrared and thermocouple temperature on backlit display
  • Last reading Hold (20 seconds)
  • High and low temperature alarm
  • Data storage and review (99 data sets)
  • Tripod mount
  • 12 or 24 hour clock
  • USB 2.0 computer interface cable
  • FlukeView® Forms documenting software
  • Two-year warranty

Features & Specification

More Information
Key Features Probe Input, Data Logging, Temperature Range 500c+, Recommended Distance Of 3.0m+, PC Interface, Contact Probe Included, Adjustable Emissivity
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Fluke Testers
Short Description
Infrared Measurements
Infrared temperature range -30 °C to 900 °C (-22 °F to 1652 °F)
IR accuracy (calibration geometry with ambient temperature 23 °C ± 2 °C) ≥ 0°C: ± 1°C or ± 1 % of the reading, whichever is greater
≥-10 °C to <0 °C: ±2°C
<-10°C: ±3 °C

≥32°F: ± 2°F or ± 1% of the reading, whichever is greater
≥14 °F to <32 °F: ±4 °F
<14 °F: ±6 °F
IR repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater
Display resolution 0.1 °C / 0.1 °F
Distance: Spot 60:1 (calculated at 90 % energy)
Minimum spot size 19 mm
Laser sighting Offset dual laser, output <1 mW
Spectral response 8 μm to 14 μm
Response time (95 %) <500 ms
Emissivity Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials

 

Contact Measurements
K-type thermocouple input temperature range -270 °C to 1372 °C (-454 °F to 2501 °F)
K-type thermocouple input accuracy (with ambient temperature 23 °C ± 2 °C) <-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
K-type thermocouple 0.1 °C/0.1 °F
K-type thermocouple repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater

 

Measurement Options
Hi/Low alarms Audible and two-color visual
Min/Max/Avg/Dif Yes
Switchable celsius and fahrenheit Yes
Backlight Two levels, normal and extra bright for darker environments
Probe input K-type thermocouple simultaneous display of probe and IR temperature
Trigger lock Yes
Data storage 99 points
Display Dot matrix 98 x 96 pixels with function menus
Communication USB 2.0

 

K-Type Thermocouple Specifications
Measurement range (bead probe) -40 °C to 260 °C (-40 °F to 500 °F)
Accuracy ±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F)
Cable length 1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector and bead termination

 

General Specifications
Safety and compliance IEC 60825-1 
FDA Laser Class II 
IEC 61326-1 
CE Complaint 
CMC # 01120009
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Relative humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating altitude 2000 meters above mean sea level
Weight 0.322 kg (0.7099 lb)
Power 2 AA batteries
Battery life 8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on)

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA