Fluke 1555C 10kV Insulation Resistance Tester Kit

£4,678.80

Electrical specifications

The tester’s accuracy is specified for one year after calibration at operating temperatures of 0 °C to 35 °C. For operating temperatures outside the range (-20 °C to 0 °C and 35 °C to 50 °C), add ± .25 % per °C, except on the 20 % bands add ± 1 % per °C. 

Insulation resistance measurement

Test voltage (dc)

Range (dc)

Accuracy (± reading)

250 V < 200 kΩ
200 kΩ to 5 GΩ
5 GΩ to 50 GΩ
> 50 GΩ
unspecified 
5 % 
20 % 
unspecified
500 V   200 kΩ
200 kΩ to 10 GΩ
10 GΩ to 100 GΩ
> 100 GΩ
unspecified 
5 %
20 % 
unspecified
1000 V  < 200 kΩ
200 kΩ to 20 GΩ
20 GΩ to 200 GΩ
> 200 GΩ
unspecified
5 %
20 %
unspecified
2500 V  < 200 kΩ
200 kΩ to 50 GΩ
50 GΩ to 500 GΩ
> 500 GΩ
unspecified
5 %
20 %
unspecified
5000 V  < 200 kΩ
200 kΩ to 100 GΩ
100 GΩ to 1 TΩ
> 1 TΩ
unspecified
5 %
20 %
unspecified
10000 V
(1555 only)
 < 200 kΩ
200 kΩ to 200 GΩ
10 GΩ to 2 TΩ
> 2 TΩ
unspecified
5 %
20 %
unspecified
Bar graph range 0 to 1 TΩ 
Insulation test voltage accuracy -0 %, +10 % at 1 mA load current
Induced ac mains current rejection 2 mA maximum

Charging rate for capacitive load

5 seconds per μF 

Discharge rate for capacitive load

1.5 s/μF

Leakage current measurement

Range: 1 nA to 2 mA

Accuracy: ± (5 % + 2 nA)

Capacitance measurement

Range: 0.01 uF to 15.00 μF

Accuracy: ± (15 % rdg + 0.03 μF)

Timer

Range: 0 to 99 minutes

Resolution Accuracy: 
Setting: 1 minute 
Indication: 1 second

Live circuit warning

Warning Range: 
30 V to 660 V ac/dc, 
50/60 Hz

Voltage Accuracy: 
± (15 % + 2 V)

 

General specifications

Display 75 mm x 105 mm
Power 12 V lead-acid rechargeable battery, Yuasa NP2.8-12
Charger Input (AC) Charger Input (AC) 85 V to 250 V ac 50/60 HZ 20 VA
This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only.
Dimensions 170 mm x 242 mm x 330 mm (6.7 in x 9.5 in x 13 in)
Weight 3.6 kg (7.94 lb)
Temperature (operating) -20 ºC to 50 ºC (-4 ºF to 122 ºF)
Temperature (storage) -20 ºC to 65 ºC (-4 ºF to 149 ºF)
Humidity 80 % to 31 ºC decreasing linearly to 50 % at 50 ºC
Altitude 2000 m
Enclosure sealing IP40
Input overload protection 1000 V ac
Electromagnetic compatibility EN 61326
Certifications
Safety compliance CAT III 1000 V, CAT IV 600 V
Pollution degree 2

Description

Fluke 1555C Digital insulation Resistance Tester Kit

Digital insulation testing up to 10kV


The new Fluke 1555 and redesigned Fluke 1550C insulation resistance testers, offer digital insulation testing up to 10kV, making them ideal for testing a wide range of high voltage equipment including switchgear, motors, generators and cables. 


Included Accessories:

  • Test Cables with Alligator Clips (red, black, green)
  • Infrared adapter with interface cable
  • FlukeView Forms Basic CD-ROM
  • AC Power Cord
  • Soft Carrying Case (base models only)
  • English Manual
  • Users Manual on CD-ROM
  • Quick Reference Card
  • Software License Agreement
  • Registration Card
  • FlukeView Forms Installation Guide
  • USB-IR Cable Installation Guide
  • IP67 Hard Case (kit only)
  • Certificate of Calibration (kit only)
  • Ruggedized Alligator Clips (kit and 1555 only)

Fluke insulation testers can now conduct the entire range of test voltages specified in IEEE 43-2000 with a best in class, 3 year warrenty and CAT IV 600 V safety rating. With measurement storage and PC interface, the 1555 and 1550C and perfect tools for pre-ventative or predictive maintenance programs designed to identify potential equipment failures before they occur. 


Features and benefits:

  • Test voltages up to 10 kV provides solutions for all applications
  • CAT III 1000 V, CAT IV 600 V safety rating
  • Voltage breakdown detection alerts the user that voltage is present and gives the voltage reading up to 600 V ac or dc for increased user safety
  • Selectable test voltages in 50 V steps from 250 to 1000 V, and 100 V steps above 1000 V
  • Measurements can be stored in up to 99 memory locations, with each location assigned a unique, user defined, label for easy recall
  • Long battery life gives the user over 750 tests between charges
  • Automatic calculation of Dielectric Absorption (DAR) and Polarization Index (PI) with no additional setup
  • Guard system eliminates the effect of surface leakage current on high-resistance measurements
  • Large digital/analog LCD for easy viewing
  • Capacitance and leakage current measurement
  • Ramp function for breakdown testing
  • Resistance measurements up to 2TΩ
  • Timer settings up to 99 minutes for timed tests
  • 3-year warranty

Features & Specification

More Information
Key Features Manufacturer Certificate Included, 10kV Tester, Internal memory, Timed Tests, Capacitance and leakage measurements, Robust and portable, IP67 Hard Case, Heavy Duty Alligator Clips
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Fluke Testers
Short Description

Electrical specifications

The tester’s accuracy is specified for one year after calibration at operating temperatures of 0 °C to 35 °C. For operating temperatures outside the range (-20 °C to 0 °C and 35 °C to 50 °C), add ± .25 % per °C, except on the 20 % bands add ± 1 % per °C. 

Insulation resistance measurement

Test voltage (dc)

Range (dc)

Accuracy (± reading)

250 V < 200 kΩ
200 kΩ to 5 GΩ
5 GΩ to 50 GΩ
> 50 GΩ
unspecified 
5 % 
20 % 
unspecified
500 V   200 kΩ
200 kΩ to 10 GΩ
10 GΩ to 100 GΩ
> 100 GΩ
unspecified 
5 %
20 % 
unspecified
1000 V  < 200 kΩ
200 kΩ to 20 GΩ
20 GΩ to 200 GΩ
> 200 GΩ
unspecified
5 %
20 %
unspecified
2500 V  < 200 kΩ
200 kΩ to 50 GΩ
50 GΩ to 500 GΩ
> 500 GΩ
unspecified
5 %
20 %
unspecified
5000 V  < 200 kΩ
200 kΩ to 100 GΩ
100 GΩ to 1 TΩ
> 1 TΩ
unspecified
5 %
20 %
unspecified
10000 V
(1555 only)
 < 200 kΩ
200 kΩ to 200 GΩ
10 GΩ to 2 TΩ
> 2 TΩ
unspecified
5 %
20 %
unspecified
Bar graph range 0 to 1 TΩ 
Insulation test voltage accuracy -0 %, +10 % at 1 mA load current
Induced ac mains current rejection 2 mA maximum

Charging rate for capacitive load

5 seconds per μF 

Discharge rate for capacitive load

1.5 s/μF

Leakage current measurement

Range: 1 nA to 2 mA

Accuracy: ± (5 % + 2 nA)

Capacitance measurement

Range: 0.01 uF to 15.00 μF

Accuracy: ± (15 % rdg + 0.03 μF)

Timer

Range: 0 to 99 minutes

Resolution Accuracy: 
Setting: 1 minute 
Indication: 1 second

Live circuit warning

Warning Range: 
30 V to 660 V ac/dc, 
50/60 Hz

Voltage Accuracy: 
± (15 % + 2 V)

 

General specifications

Display 75 mm x 105 mm
Power 12 V lead-acid rechargeable battery, Yuasa NP2.8-12
Charger Input (AC) Charger Input (AC) 85 V to 250 V ac 50/60 HZ 20 VA
This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only.
Dimensions 170 mm x 242 mm x 330 mm (6.7 in x 9.5 in x 13 in)
Weight 3.6 kg (7.94 lb)
Temperature (operating) -20 ºC to 50 ºC (-4 ºF to 122 ºF)
Temperature (storage) -20 ºC to 65 ºC (-4 ºF to 149 ºF)
Humidity 80 % to 31 ºC decreasing linearly to 50 % at 50 ºC
Altitude 2000 m
Enclosure sealing IP40
Input overload protection 1000 V ac
Electromagnetic compatibility EN 61326
Certifications
Safety compliance CAT III 1000 V, CAT IV 600 V
Pollution degree 2

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA