Fluke 123B 20 MHz Industrial Scope Meter
Product Comparison Chart | Fluke 123B | Fluke 124B | Fluke 125B |
Functions | |||
Full Function Dual Input Scope and Meter | • | • | • |
Oscilloscope Bandwidth MHz | 20 | 40 | 40 |
Meter and Scope Recorder | • | • | • |
Scope Cursor Measurements | • | • | |
Power and Harmonics Measurements | • | ||
BusHealth | • | ||
Included Accessories | |||
10:1 voltage probe | • | • | |
i400S AC current clamp | • |
Oscilloscope mode | ||
Vertical | ||
Frequency response - dc coupled | Without probes and test leads (with BB120) | 123B: dc to 20 MHz (-3 dB) 124B and 125B: dc to 40 MHz (-3 dB) |
With STL120-IV 1:1 shielded test leads | DC to 12.5 MHz (-3 dB) / dc to 20 MHz (-6 dB) | |
With VP41 10:1 Probe | 123B: dc to 20MHz (-3 dB) 124B and 125B: dc to 40 MHz (-3 dB) |
|
Frequency response - ac coupled (lf roll off) | Without probes and test leads | <10 Hz (-3 dB) |
With STL120-IV 1:1 shielded test leads | <10 Hz (-3 dB) | |
With VP41 10:1 Probe | <10 Hz (-3 dB) | |
Rise time, excluding probes, test leads |
123B <17.5 ns 124B and 125B <8.75 ns |
|
Input impedance | Without probes and test leads | 1 MΩ//20 pF |
With BB120 | 1 MΩ//24 pF | |
With STL120-IV 1:1 shielded test leads | 1 MΩ//230 pF | |
With VP41 10:1 Probe | 5 MΩ//15.5 pF | |
Sensitivity | 5 mV to 200 V/div | |
Analog bandwidth limiter | 10 kHz | |
Display modes | A, -A, B, -B | |
Max. input voltage A and B | Direct, with test leads, or with VP41 Probe | 600 Vrms Cat IV, 750 Vrms maximum voltage. |
With BB120 | 600 Vrms | |
Max. floating voltage, from any terminal to ground | 600 Vrms Cat IV, 750 Vrms up to 400Hz | |
Horizontal | ||
Scope modes | Normal, Single, Roll | |
Ranges (normal) | Equivalent sampling | 123B: 20 ns to 500 ns/div, |
124B and 125B: 10 ns to 500 ns/div | ||
Real time sampling | 1 μs to 5 s/div | |
Single (real time) | 1 μs to 5 s/div | |
Roll (real time) | 1s to 60 s/div | |
Sampling rate (for both channels simultaneously) | Equivalent sampling (repetitive signals) | Up to 4 GS/s |
Real time sampling 1 μs to 60 s/div | 40 MS/s |
Trigger | ||
Screen update | Free run, on trigger | |
Source | A, B | |
Sensitivity A and B | @ DC to 5 MHz | 0.5 divisions or 5 mV |
@ 40 MHz | 123B: 4 divisions | |
124B and 125B: 1.5 divisions | ||
@ 60 MHz | 123B: N/A | |
124B and 125B: 4 divisions | ||
Slope | Positive, negative |
Advanced scope functions | ||
Display modes | Normal | Captures up to 25 ns glitches and displays analog-like persistence waveform |
Smooth | Suppresses noise from a waveform | |
Glitch off | Does not capture glitches between samples | |
Envelope | Records and displays the minimum and maximum of waveforms over time | |
Auto set (Connect-and-View™) | Continuous fully automatic adjustments of amplitude, time base, trigger levels, trigger gap, and hold-off. Manual override by user adjustment of amplitude, time base, or trigger level. |
Dual input meter | ||
The accuracy of all measurements is within ±(% of reading + number of counts) from 18 °C to 28 °C. | ||
Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1%. More than one waveform period must be visible on the screen. |
Input A and input B | ||
DC voltage (VDC) | ||
Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V | |
Accuracy | ±(0.5% +5 counts) | |
Common mode rejection (CMRR) | >100 dB @ dc, >60 dB @ 50, 60, or 400 Hz | |
Full scale reading | 5000 counts | |
True-rms voltages (V ac and V ac+dc) | ||
Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V | |
Accuracy for 5% to 100% of range (DC coupled) | DC to 60 Hz (V ac+dc) | ±(1% +10 counts) |
1 Hz to 60 Hz (V ac) | ±(1% +10 counts) | |
Accuracy for 5% to 100% of range (AC or dc coupled) | 60 Hz to 20 kHz | ±(2.5% +15 counts) |
DC rejection (only VAC) | >50 dB | |
Common mode rejection (CMRR) | >100 dB @ dc | |
>60 dB @ 50, 60, or 400 Hz | ||
Full scale reading | 5000 counts, reading is independent of any signal crest factor. | |
Peak | ||
Modes | Max peak, Min peak, or pk-to-pk | |
Ranges | 500 mV, 5 V, 50 V, 500 V, 2200 V | |
Accuracy | Accuracy Max peak or Min peak | 5% of full scale |
Accuracy Peak-to-Peak | 10% of full scale | |
Full scale reading | 500 counts | |
Frequency (Hz) | ||
Ranges | 123B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz | |
124B and 125B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz, and 70 MHz | ||
Frequency range | 15 Hz (1 Hz) to 50 MHz in continuous autoset | |
Accuracy @1 Hz to 1 MHz | ±(0.5% +2 counts) | |
Full scale reading | 10,000 counts | |
RPM | ||
Max reading | 50.00 kRPM | |
Accuracy | ±(0.5% +2 counts) | |
Duty cycle (PULSE) | ||
Range | 2% to 98% | |
Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset | |
Pulse width (PULSE) | ||
Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset | |
Full scale reading | 1000 counts | |
Amperes (AMP) | ||
With current clamp | Ranges | Same as V dc, V ac, V ac+dc, or PEAK |
Scale factors | 0.1 mV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400 mV/A, 1 V/A, 10 mV/mA | |
Accuracy | Same as V dc, V ac, V ac+dc, or PEAK (add current clamp uncertainty) | |
Temperature (TEMP) with optional temperature probe | ||
Range | 200 °C/div (200 °F/div) | |
Scale factor | 1 mV/°C and 1 mV/°F | |
Accuracy | As V dc (add temp. probe uncertainty) | |
Decibel (dB) | ||
0 dBV | 1 V | |
0 dBm (600 Ω /50 Ω) | 1 mW referenced to 600 Ω or 50 Ω | |
dB on | V dc, V ac, or Vac+dc | |
Full scale reading | 1000 counts | |
Crest factor (CREST) | ||
Range | 1 to 10 | |
Full scale reading | 90 Counts | |
Phase | ||
Modes | A to B, B to A | |
Range | 0 to 359 degrees | |
Resolution | 1 degree | |
Power (125B only) | ||
Configurations | 1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only) | |
Power factor (PF) | Ratio between watts and VA range - 0.00 to 1.00 | |
Watt | RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes) | |
Full scale reading | 999 counts | |
VA | Vrms x Arms | |
Full scale reading | 999 counts | |
VA reactive (var) | √((VA)2-W2) | |
Full scale reading | 999 counts | |
Vpwm | ||
Purpose | To measure on pulse width modulated signals, like motor drive inverter outputs | |
Principle | Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency | |
Accuracy | As Vrms for sinewave signals |
Input A to common | ||
Ohm (Ω) | ||
Ranges | 123B and 124B | 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ |
125B | 50 Ω, 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ | |
Accuracy | ±(0.6% + 5 counts) 50 Ω ±(2% + 20 counts) | |
Full scale reading | 50 Ω to 5 MΩ - 5000 counts, 30 MΩ - 3000 counts | |
Measurement current | 0.5 mA to 50 nA, decreases with increasing ranges | |
Open circuit voltage | <4 V | |
Continuity (Cont) | ||
Beep | <(30 Ω ±5 Ω) in 50 Ω range | |
Measurement current | 0.5 mA | |
Detection of shorts of | ≥1 ms | |
Diode | ||
Measurement voltage | @0.5 mA | >2.8 V |
@open circuit | <4 V | |
Measurement current | 0.5 mA | |
Polarity | + on input A, - on COM | |
Capacitance (CAP) | ||
Ranges | 50 nF, 500 nF, 5 μF, 50 μF, 500 μF | |
Full scale reading | 5000 counts | |
Measurement current | 500 nA to 0.5 mA, increases with increasing ranges |
Advanced meter functions | ||
Zero Set | Set actual value to reference | |
AutoHold (on A) | Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals. | |
Fixed decimal point | Activated by using attenuation keys |
Cursor Readout (124B and 125B) | ||
Sources | A, B | |
Single vertical line | Average, min and max readout | |
Average, min, max and time from start of readout (in ROLL mode; instrument in HOLD) | ||
Min, max and time from start of readout (in RECORDER mode; instrument in HOLD) | ||
Harmonics values in POWER QUALITY mode. | ||
Dual vertical lines | Peak-peak, time distance and reciprocal time distance readout | |
Average, min, max and time distance readout (in ROLL mode; instrument in HOLD) | ||
Dual horizontal lines | High, low and peak-peak readout | |
Rise or fall time | Transition time, 0%-level and 100%-level readout (manual or auto leveling; auto leveling only possible in single channel mode) |
|
Accuracy | As oscilloscope accuracy |
Recorder | ||
The recorder captures meter readings in Meter Recorder mode or continuously captures waveform samples in Scope Recorder mode. The information is stored on internal memory or on optional SD card (with the 125B or 124B). | ||
The results are displayed as Chart recorder display that plots a graph of min and max values of Meter measurements over time or as a waveform recorder display that plots all the captured samples. | ||
Meter readings | ||
Measurement Speed | Maximum 2 measurement/s | |
Record Size (min, max, average) | 2 M readings for 1 channel | |
Recorded Time Span | 2 weeks | |
Maximum number of events | 1024 | |
Waveform record | ||
Maximum sample rate | 400 K sample/s | |
Size Internal memory | 400 M samples Recorded Time | |
Span internal memory | 15 minutes at 500 μs/div 11 hours at 20 ms/div |
|
Record Size SD card | 1.5 G samples | |
Recorded Time Span SD card | 11 hours at 500 μs/div 14 days at 20 ms/div |
|
Maximum number of events | 64 |
Power Quality (125B only) | ||
Readings | Watt, VA, var, PF, DPF, Hz | |
Watt, VA, var ranges (auto) | 250 W to 250 MW, 625 MW, 1.56 GW | |
When selected: total (% r) | ±(2% + 6 counts) | |
When selected: fundamental (% f) | ±(4% + 4 counts) | |
DPF | 0.00 to 1.00 | |
PF | 0.00 to 1.00, ±0.04 | |
Frequency range | 10.0 Hz to 15.0 kHz 40.0 Hz to 70.0 Hz |
|
Number of Harmonics | DC to 51 | |
Readings / Cursor readings (fundamental 40 Hz to 70 Hz) |
Vrms / Arms /Watt | Each harmonic from fundamental maybe selected for individual readings |
Includes frequency of fundamental, phase Angle and K-factor (in Amp and Watt) |
BusHealth tester (Fluke 125B only) | ||
Type | Subtype | Protocol |
AS-i | NEN-EN50295 | |
CAN | ISO-11898 | |
Interbus S | RS-422 | EIA-422 |
Modbus | RS-232 | RS-232/EIA-232 |
RS-485 | RS-485/EIA-485 | |
Foundation Fieldbus | H1 | 61158 type 1, 31.25 kBit |
Profibus | DP | EIA-485 |
PA | 61158 type 1 |
Miscellaneous | ||
Display | Type | 5.7-inch color active matrix TFT |
Resolution | 640 x 480 pixels | |
Waveform Display | Vertical | 10 div of 40 pixels |
Horizontal | 12 div of 40 pixels | |
Power | External | Via Power Adapter BC430 |
Input voltage | 10 V DC to 21 V DC | |
Power consumption | 5 W typical | |
Input connector | 5 mm jack | |
Internal | Via Battery Pack BP290 | |
Battery power | Rechargeable Li-Ion 10.8 V | |
Operating time | 7 hours with 50% backlight brightness | |
Charging time | 4 hours with test tool off, 7 hours with test tool on | |
Allowable ambient temp | 0 to 40 °C (32 to 104 °F) during charging | |
Memory | Internal memory can store 20 data sets (screen waveform and setup) | Micro SD card slot with optional SD card (max size of 32 GB) |
Mechanical | Size | 259 x 132 x 55 mm (10.2 x 5.2 x 2.15 in) |
Weight | 1.4 kg (3.2 lb) including battery pack | |
Interface | Optically isolated | Transfer screen copies (bitmaps), settings and data |
USB to PC/laptop | OC4USB optically isolated USB adapter/cable, (optional), using FlukeView® software for Windows®. | |
Optional WiFi adapter | Fast transfer of screen copies (bitmaps), settings and data to PC/laptop, tablet, smartphone, etc. A USB port is provided for attaching the WiFi dongle. Do not use the USB port with a cable for safety reasons. |
Environmental | ||
Environmental | MIL-PRF-28800F, Class 2 | |
Temperature | Battery Operation | 0 to 40 °C (32 to 104 °F) |
Power Adapter Operation | 0 to 50 °C (32 to 122 °F) | |
Storage | -20 to 60 °C (-4 to 140 °F) | |
Humidity (Operating) | @ 0 to 10 °C (32 to 50 °F) | Non-condensing |
@ 10 to 30 °C (50 to 86 °F) | 95% | |
@ 30 to 40 °C (86 to 104 °F) | 75% | |
@ 40 to 50 °C (104 to 122 °F) | 45% | |
Storage | @ -20 to 60 °C (-4 to 140 °F) | Non-condensing |
Altitude | Operating at 3 km (10,000 feet) | CAT III 600 V |
Operating at 2 km (6,600 feet) | CAT IV 600 V | |
Storage | 12 km (40,000 feet) | |
EMC electromagnetic compatibility | International | IEC 61326-1: Industrial, CISPR 11: Group 1, Class A |
Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) | |
USA (FCC) | 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. | |
Wireless radio with adapter | Frequency range | 2412 MHz to 2462 MHz |
Output power | <100 mW | |
Enclosure protection | IP51, ref: EN/IEC60529 | |
Safety | General | IEC 61010-1: Pollution Degree 2 |
Measurement | IEC 61010-2-033: CAT IV 600 V/CAT III 750 V | |
Max. input voltage input A and B | Direct on input or with leads | 600 Vrms CAT IV for derating |
With Banana-to-BNC Adapter BB120 | 600 Vrms for derating | |
Max. floating voltage from any terminal to ground | 600 Vrms Cat IV, 750 Vrms up to 400 Hz |
Description
Fluke 120B Series Industrial ScopeMeter® Hand-Held Oscilloscopes
Simplified testing, more insight and faster electro-mechanical troubleshooting
The Fluke ScopeMeter® 120B Series is a rugged oscilloscope for industrial electrical and electro-mechanical equipment troubleshooting and maintenance.
The compact ScopeMeter® 120B Series, is the rugged oscilloscope solution for industrial electrical and electro-mechanical equipment troubleshooting and maintenance applications. It’s a truly integrated test tool, with oscilloscope, multimeter and high-speed recorder in one easy-to-use instrument. The ScopeMeter 120B Series also integrates with Fluke Connect® mobile app and FlukeView® for ScopeMeter software to enable further collaboration, data analysis and archiving of critical test information.
Fluke 123B Scope Meter Includes:
- Fluke 123B Hand-Held Oscilloscope
- Shielded Test Leads with Black Ground Leads
- Test Lead Black (for Grounding)
- Hook Clips (red, blue)
- Banana-to-BNC Adapters (black, x1)
- USB Angled Adapter
- WiFi USB Adapter*
- Switch Mode Power Supply, Adapter/Battery Charger
- Rechargeable Li-ion Battery Pack
Fluke 123B/S Scope Meter Includes:
- All of the above +
- Soft Carrying Case
- Magnetic Hanger
- FlukeView® ScopeMeter® Software for Windows®
- Screen Protector
The 120B Series Industrial ScopeMeter test tools include innovative functions designed to help technicians troubleshoot faster and get the answers they need to keep their systems up and running.
- Dual-input digital oscilloscope and multimeter
- 40 MHz or 20 MHz oscilloscope bandwidth
- Two 5,000-count true-rms digital multimeters
- Connect-and-View™ trigger simplicity for hands-off operation
- IntellaSet™ technology automatically and intelligently adjusts numerical readout based on the measured signal
- Dual-input waveform and meter reading recorder for trending data over extended periods
- Recorder Event Detect captures elusive intermittent signals on repetitive waveforms up to 4 kHz
- Shielded test leads for oscilloscope, resistance and continuity measurements
- Resistance, continuity, diode and capacitance meter measurements
- Power measurements (W, VA, VAR, PF, DPF, Hz)
- Voltage, current and power harmonics
- Check Industrial networks with BusHealth physical layer tests against defined reference levels
- Save or recall data and instrument setups
- Store instrument setups defined by a test sequence for routine maintenance or most often used test procedures.
- External optically isolated USB interface to transfer, archive and analyze scope or meter data
- Optional WiFi adapter connected to internal USB port to wirelessly transfer information to the PC, laptop or Fluke Connect® mobile app*
- FlukeView® ScopeMeter® Software for Windows®
- Rugged design to withstand 3g Vibration, 30g shock, and rated IP51 according to EN/IEC60529
- Highest safety rating in the industry: safety rated for CAT IV 600 V
- Li-Ion rechargeable battery, seven-hours operation (with four-hour charge time)
Connect-and-View™ triggering foran instant, stable display
Oscilloscope users know how difficult triggering can be. Using the wrong settings can lead to unstable waveform captures, and sometimes the wrong measurement data. Fluke's unique Connect-and-View™ triggering technology recognizes signal patterns, and automatically sets up the correct triggering to provide a stable, reliable and repeatable display. Connect-and-View™ triggering is designed to work with virtually any signal, including motor drives and control signals - without adjusting parameters, or even touching a button. Signal changes are instantly recognized and settings are automatically adjusted, providing a stable display even when measuring multiple test points in quick succession.
IntellaSet™/AutoReading
The Auto Readings function with Fluke IntellaSet™ technology uses proprietary algorithms to intelligently analyze the measured waveform and automatically displays the most appropriate numerical measurements on screen, so you can get the data you need easier than ever before. As an example, when the measured waveform is a line voltage signal, the Vrms and Hz readings are automatically displayed, whereas if the measured waveform is a square wave, the Vpeak-peak and Hz readings are automatically displayed. Using IntellaSet™ technology in conjunction with Connect-and-View™ automatic triggering you can be sure you're seeing not only the correct waveform, but the appropriate numerical reading as well. All without touching a button.
Industrial equipment needs a reliable power supply to operate properly, use the dual input to obtain key power measurements.
For single phase or 3-phase balanced systems, the dual inputs of the Industrial ScopeMeter® 120B Series can measure ac+dc rms voltage on channel A and ac+dc rms current on channel B. The Fluke 125B can then calculate; frequency, phase angle, active power (kW), reactive power (VA or var), power factor (PF) or displacement power factor (DPF) and can also calculate the power values for a 3-phase system where all phases have equal voltage and currents. This applies to both balanced system and resistive loads.
Harmonics measurements
Harmonics are periodic distortions of voltage, current, or power sine waves. Harmonics in power distribution systems are often caused by non-linear loads such as switched mode dc power supplies and adjustable speed motor drives. Harmonics can cause transformers, conductors, and motors to overheat. In the Harmonics function, the Test Tool measures harmonics to the 51st. Related data such as dc components, THD (Total Harmonic Distortion), and K factor are measured to provide a complete insight in to the electrical state of health of your loads.
Use the comprehensive recorder modes to help find intermittent faults with ease
The toughest faults to find are those that happen only once in a while - intermittent events. They can be caused by bad connections, dust, dirt, corrosion or simply broken wiring or connectors. Other factors, like line outages and sags or the starting and stopping of a motor, can also cause intermittent events resulting in equipment shutdowns. When these events happen, you may not be around to see it. But, your Fluke ScopeMeter® Test Tool will.
You can either plot the minimum and maximum peak measurement values or record the waveform trace. And, with expandable micro SD memory, recording sessions can be done for up to 14 days. This recorder is even more powerful with the addition of Recorder Event Detect, which makes detecting and logging intermittent faults easier than ever. Just set a threshold on a meter reading or scope trace and deviations are tagged as unique events. You no longer need to search through masses of data to pinpoint faults, and can quickly step from one tagged event to the next, while still having access to the full data set
Industrial BusHealth Test verifies electrical signal quality on industrial buses
BusHealth Test analyzes the electrical signals on the industrial bus or network and gives a clear "Good, "Weak or "Bad indication mark for each of the relevant parameters, presented next to the actual measurement value. Measured values are compared to standard values based on the selected bus types (CAN-bus, Profi-bus, Foundation Field, RS-232 and many more), or, unique reference values can be set if different tolerances are required.
The Fluke 125B can validate the quality of the electrical signals as soon as they are passed along the network, without looking at the data content. Additionally, the 125B checks the signal levels and speed, transition times and distortion, and compares these to the appropriate standards to help you find errors such as improper cable connections, bad contacts, incorrect grounding or improper terminators.
One test lead to measure multiple electrical parameters
High frequency waveform, meter, capacitance and resistance measurements as well as continuity checks are all covered by single set of shielded test leads. No time is wasted finding or swapping leads.
FlukeView® ScopeMeter® Software for Windows®
Get more out of your ScopeMeter 120B with FlukeView® Software:
- Store instrument's screen copies on the PC, in color
- Copy screen images into your reports and documentation
- Capture and store waveform data from your ScopeMeter test tool on your PC
- Create and archive waveform references for easy comparison
- Copy waveform data into your spreadsheet for detailed analysis
- Use cursors for parameter measurement
- Add user text to instrument setups and send them to the instrument for operator reference and instructions
Fluke Connect mobile app compatibility
Automated industrial machinery is harder than ever to trouble shoot. It's not enough to just know where you have to test, you also have to know what to look for and that can be hard without baseline measurement data or access to subject matter experts. The Fluke Connect® Assets wireless system of software and wireless test tools enables technicians to reduce maintenance costs and increase uptime with accurate equipment records and maintenance data that is easy to interpret, and share.
Download the Fluke Connect® app
Compare and contrast test point measurement data and trends so you can better understand signal characteristics and changes over time. And, by storing maintenance data on the Fluke Cloud™ you can enable team members to access it from wherever and whenever they need to so you can get advice or approvals in the field and get your systems up and running faster than ever before.
Product Comparison Chart |
Fluke 123B |
Fluke 124B |
Fluke 125B |
Full Function Dual Input Scope and Meter |
* |
* |
* |
Oscilloscope Bandwidth MHz |
20 |
40 |
40 |
Meter and Scope Recorder |
* |
* |
* |
Scope Cursor Measurements |
|
* |
* |
Power and Harmonics Measurements |
|
|
* |
BusHealth |
|
|
* |
10:1 voltage probe |
|
* |
* |
i400S AC current clamp |
|
|
* |
Features & Specification
Key Features | Large Backlit Display, Colour Display, 3 instruments in 1, Rechargeable Batteries, 2 channel 20MHz | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Product Stock Status | Next Day Delivery (subject to availability) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Manufacturer | Fluke Testers | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Short Description |
|
Delivery
Updated January 2024
Please note, our cut-off time is 2:00pm
Orders requiring calibration
Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.
Bank/Public Holidays
Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scopeof our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.
*UK Mainland (Revised March 2021)
Service | Price |
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT | FREE |
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT | £4.95 |
Next WORKING Day (excludes Saturdays, Sundays and public holidays) – UK Mainland only - Zones A & B - ANY order value - Order by 2:00pm | £7.75 |
Saturday ANYTIME Delivery - Order by 2:00pm | £9.95 |
Delivery Surcharges
Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:
- Zones A & B - Mainland delivery rates apply.
- Zone C - Scottish Highlands (Charges are calculated at the checkout).
- Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
- Zone E - Northern Ireland (Charges are calculated at the checkout).
- Zone F - Republic of Ireland (Charges are calculated at the checkout).
Offshore Services
All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.
International delivery
International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk
Export
All export inquiries should be directed to the export division before placement of order.
If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA