Chauvin Arnoux Metrix Scopix IV Oscilloscope OX9062
Chauvin Arnoux Metrix Scopix IV Oscilloscopes
Type of display | 7‘’ WVGA colour TFT LCD touch screen, 800x480 – LED backlighting (adjustable standby mode) |
Different display mode | 2,500 real acquisition points on screen - Vectors with interpolation |
Display of curves on screen | 4 curves + 4 references – Split Screen & Full-Screen modes |
Screen commands | Touchscreen – ANDROID-type icons and graphical commands – customizable channel colours |
Choice of language | 15 complete languages, menus & online help |
Oscilloscope Mode | |
Input impedance | 1 MΩ ± 0.5% , approx. 12 pF |
Maximum input voltage | 600 V / CAT III (1,000V per Probix) – from 50 to 400 Hz – Probix safety connectors |
Vertical sensitivity | 16 calibres from 2.5 mV to 200 V/div and up to 156 µV/div in vertical zoom mode (12-bit converter) – Accuracy ± 2% |
Vertical zoom | "One Click Winzoom" mode (12-bit converter and direct graphical zoom on screen) – x 16 max. |
Probe factor (non-Probix) | 1 / 10 / 100 / 1,000 or any scaling – definition of measurement unit |
Horizontal Deflection | |
Sweep speed | 35 calibres from 1 ns/div to 200 s/div., accuracy ± [50 ppm + 500 ps] – Roll mode from 100 ms to 200 s/div |
Horizontal zoom | "One Click Winzoom" system (direct graphical zoom on screen) x 1 to x 5 or x 100 – storage 100 kpts/channel |
Triggering | |
Mode | On all the channels: automatic, triggered, one-shot, auto level 50% |
Type | Edge, pulse width (16 ns-20 s), delay (48 ns to 20 s), counting (3 to 16,384 events) Continuous adjustment of Trigger position |
Coupling | AC, DC GND, HFR, LFR, noise – Level and Hold-Off adjustable from 64 ns to 15 s |
Sensitivity | ≤ 1.2 division p-p up to 300 MHz |
Digital Storage | |
Maximum sampling rate | 2.5 GS/s in one-shot mode on each channel (100 GS/s max. in ETS mode) |
Vertical resolution | 12 bits (vertical resolution 0.025 %) |
Memory depth | 100 kpts per channel and file viewer in the manager |
User storage File management |
Internal = 1 GB to store the files: trace, text, configuration, math functions, System memory: .pdf print files, .png image files + high-capacity removable µSD-Card: SD 2 GB, SDHC 4-32 GB and SDXC > 32 GB |
GLITCH mode | Duration ≤ 2 ns – 500,000 Min/Max pairs |
Display modes | Envelope, vector, accumulation-, averaging (factors 2 to 64) – XY (vector) and Y(f)=FFT |
Other Functions | |
AUTOSET | Complete in under 5 s, with recognition of the channels – Frequency > 30 Hz |
FFT analyser & MATH functions | 2,500-point FFT (Lin or Log) with measurement cursors – Functions + , - , x , / and mathematical function editor |
Cursors | 2 or 3 cursors: simultaneous V and T with AUTO measurement: T1, T2, Dt, 1/Dt, dBV, Ph |
Automatic measurements | Simultaneously with waveform, 20 automatic measurements per channel and on the 4 channels simultaneously with scroll |
Multimeter Mode | |
General specifications | 2 or 4 channels – 8,000 cts min/max/frequency/relative – TRMS – Time/date-stamped graphical recording in logger mode |
AC, DC and AC + DC voltages | 600 mV to 600 VRMS, 800 mV to 800 VDC – VDC accuracy +/- (0.5 % + 25 D) – 200 kHz bandwidth |
Resistance | 80 Ω to 32 MΩ – accuracy 0.5%R+ 25D – Quick continuity test < 10 ms |
Other measurements | Temperature (HX0035 = KTC, HX0036 = Pt100) / Capacitance 5nF to 5mF / Frequency 200 kHz / Diode test 3.3 V |
Single and three-phase power | Active, Reactive and Apparent power values plus Power Factor simultaneously with the U & I measurements |
Harmonic Analyser Mode | |
Multi-channel analysis | 2 or 4 (depending on model), 63 orders, fundamental frequency 40 to 450 Hz in auto or manual mode |
Simultaneous measurements | Total Vrms, THD and selected order (% fundamental, phase, frequency, Vrms) |
Logger Mode | |
Acquisition | Duration: 20,000 s – Interval: 0.2 s – Files: 100,000 measurements |
General Specifications | |
Configuration memories | Not limited according to device - variable file sizes |
Printing | Network printing via Ethernet/Wifi in .png format |
PC communication – software link | Ethernet (100 baseT), WiFi-USB (device, 12 Mbs) – "ScopeNet" application software for PC |
Software |
PC: Ethernet and USB, ScopeNet (remote control, data recovery, cursors and automatic measurements) Android tablet – ScopeAdmin Fleet Administration utility |
Mains power supply |
Li-Ion rechargeable battery (6,900mAH-40 Wh) – Battery life of up to 8 hrs – Adjustable standby mode Adapter / 2-hour fast charger, universal 98-264 V / 50/60 Hz) |
Safety / EMC | Safety as per IEC 61010-2-30, 2010 – 600 V CAT III / 1000 V CAT II – EMC as per EN61326-1, 2010 |
Mechanical specifications | 292.5 x 210.6 x 66.2 mm – 2.1 kg with batteries – IP54 protection |
Description
Chauvin Arnoux Metrix Scopix IV Oscilloscopes
Chauvin Arnoux Scopix IV Oscilloscopes w/ Isolated Channels function as oscilloscopes, multimeters, harmonics analysers and data loggers. They are suitable for conducting electrical, Fieldbus and industrial maintenance applications.
The Chauvin Arnoux Scopix IV Oscilloscopes w/ Isolated Channels is available in four models which vary with regards to their bandwidths and the number of isolated channels. For further information, please see the table below.
Product Code | Product Name | Bandwidth | Bandwidth Limiter | Number of Isolated Channels |
OX9062 | Chauvin Arnoux OX9062 Scopix IV Oscilloscope | 60 MHz | 15 MHz, 1.5 MHz or 5 kHz bandwidth limiter | 2 |
OX9102 | Chauvin Arnoux OX9102 Scopix IV Oscilloscope | 100 MHz | 2 | |
OX9104 | Chauvin Arnoux OX9104 Scopix IV Oscilloscope | 100 MHz | 4 | |
OX9304 | Chauvin Arnoux OX9304 Scopix IV Oscilloscope | 300 MHz | 4 |
The Chauvin Arnoux Scopix IV Oscilloscopes Are Supplied With:
- Your Choice of Chauvin Arnoux Scopix IV Oscilloscopes w/ Isolated Channels
- 1x Carrying Bag
- 1x PA40W-2 Mains Power Pack/Charger
- 1x 2P EURO Mains Power Cable
- 1x Li-Ion battery pack
- 1x Stylus
- 1x Ethernet Cable
- 1x USB Cable
- 2x Safety Leads (red, black)
- 2 x Ø 4 mm Test Probes (red, black)
- Voltage Probes (2 or 4 depending on model)
- 1x μSD card (8 GB)
- 1x USB/ μSD adapter
- 1x hand strap
- 1x PROBIX BANANA
- 1x USB Installation Procedure for use of ScopeNet Data Export Software on CD-ROM
- 1x PDF user’s manual (>5 languages)
- 1x Start-up Guide on Paper
- 1x Safety Datasheet in 20 Languages
Key Features:
- Four modes: Oscilloscope, Multimeter, Analyser and Recorder
- Trigger functions: pulse width, counting, and delay
- IP54
- View RMS level, harmonic distortion, and harmonic frequency
- View the harmonic analysis of all the isolated functions simultaneously
- 8000-count, true-RMS digital multimeter
- AC/DC voltage
- AC/DC
- Resistance
- Continuity
- Capacitance
- Capture screenshots in PNG formats
- The front panel can be used to send documents to print and transfer and delete files in the file manager
- Recall curves and screenshots from the instrument’s memory
- Communicate with a PC, tablet, or smartphone via an Ethernet or USB connection
- An Ethernet LAN network with an integrated DHCP server may be used to activate a Wi-Fi radio link
Ergonomics
With its backlit 7" WVGA TFT colour touchscreen on a black background for easier reading, menus available in more than 13 languages and the design of the keyboard and casing, the SCOPIX IV are particularly comfortable to use.
New keypad design
Configuration and display of the measurements are simple thanks to the accesses on the front panel in one of the 5 specific areas: Utilities (brightness, full screen, screenshot), Measurements, Vertical, Horizontal, and Trigger.
Identification of the channels and parameters
Each channel and the related parameters can be identified because they have an identical colour against a black background for simpler, quicker viewing.
Adjustable transport strap
This helps to optimize the operation of the oscilloscope in your hand or on your shoulder when working in the field. A stand is also available to vary the orientation of the oscilloscope when it is placed on a bench. The oscilloscope can be left without supervision thanks to the Kensington locking system.
isswww and Chauvin Arnoux
ISSWWW has a selection of Chauvin Arnoux products, such as Microhmmeters, Earth Ground Clamp Meters, and Range Earth Testers, for efficient and accurate measuring. Chauvin Arnoux is constantly developing and manufacturing portable measurement instrumentation for electrical tradesmen, installers, industries, and administrations.
Features & Specification
Key Features | Four Models Available, IP54 Rated, 5 instruments in 1, Software Included, Touch Screen, Colour Display, Built In Recorder Function | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Product Stock Status | Next Day Delivery (subject to availability) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Manufacturer | Chauvin Arnoux | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Short Description | Chauvin Arnoux Metrix Scopix IV Oscilloscopes
|
Delivery
Updated January 2024
Please note, our cut-off time is 2:00pm
Orders requiring calibration
Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.
Bank/Public Holidays
Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scopeof our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.
*UK Mainland (Revised March 2021)
Service | Price |
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT | FREE |
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT | £4.95 |
Next WORKING Day (excludes Saturdays, Sundays and public holidays) – UK Mainland only - Zones A & B - ANY order value - Order by 2:00pm | £7.75 |
Saturday ANYTIME Delivery - Order by 2:00pm | £9.95 |
Delivery Surcharges
Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:
- Zones A & B - Mainland delivery rates apply.
- Zone C - Scottish Highlands (Charges are calculated at the checkout).
- Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
- Zone E - Northern Ireland (Charges are calculated at the checkout).
- Zone F - Republic of Ireland (Charges are calculated at the checkout).
Offshore Services
All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.
International delivery
International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk
Export
All export inquiries should be directed to the export division before placement of order.
If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA