Fluke 754 Documenting Process Calibrator HART

£8,638.80
Measurement Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV        0.02%+0.005 mV  0.03%+0.005 mV
3.00000 V  0.02%+0.00005 V  0.03%+0.00005 V
30.0000 V  0.02%+0.0005 V  0.03%+0.0005 V
300.00 V  0.05%+0.05 V  0.07%+0.05 V
Voltage AC
Range 
40 to 500 Hz     
 Resolution  1 Year 2 Years 
3.000 V  0.001 V  0.5%+0.002 V  1.0%+0.004 V
30.00 V  0.01 V  0.5%+0.02 V  1.0%+0.04 V
300.0 V  0.1 V  0.5%+0.2 V  1.0%+0.2 V
Current DC
   1 Year  2 Years
30.000 mA          0.01% + 5 uA   0.015%+7 uA
110.00 mA  0.01% + 20 uA  0.015%+30 uA
Resistance
   1 Year  2 Years
10.000 Ω             0.05% + 50 mΩ  0.07%+70 mΩ
100.00 Ω  0.05% + 50 mΩ  0.07%+70 mΩ
1.0000 kΩ   0.05% + 500 mΩ  0.07%+0.5 Ω
10.000 kΩ  0.1% + 10 Ω  0.15%+15 Ω
Frequency
   Resolution  Accuracy (2 Years)
1.00 to 110.00 Hz  0.01 Hz  0.05 Hz
110.1 to 1100.0 Hz  0.1 Hz  0.5 Hz
1.101 to 11.000 kHz   0.001 kHz  0.005 kHz
11.01 to 50.00 kHz  0.01 kHz  0.05 kHz

 

Source Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV            0.01%+0.005 mV  0.015%+0.005 mV
1.00000 V  0.01%+0.00005 V  0.015%+0.0005 V
15.0000 V  0.01%+0.0005 V  0.015%+0.0005 V
Current DC
   1 Year 2 Years 
22.000 mA (Source)  0.01%+ 0.003 mA  0.02%+ 0.003 mA
Current Sink (Simulate)  0.02% + 0.007 mA  0.04% + 0.007 mA
Resistance
   1 Year  2 Years
10.000 Ω                0.01% + 10 mΩ  0.015% + 15 mΩ
100.00 Ω  0.01% + 20 mΩ  0.015% + 30 mΩ
1.0000 kΩ  0.02% + 0.2 Ω  0.03% + 0.3 Ω
10.000 kΩ  0.02% + 3 Ω  0.03% + 5 Ω
Frequency
   2 Years
0.1 to 10.99 Hz  0.01 Hz
0.01 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1.100 to 21.999 kHz  0.002 kHz
22.000 to 50.000 kHz           0.005 kHz

 

Technical Data
Data Log Functions
Measure functions:  Voltage, current, resistance, frequency, temperature, pressure
Reading rate:  1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length:  8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Rate:  4 steps/second
Trip detect:  Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage:  Selectable, 26 V
Accuracy:  10%, 18 V minimum at 22 mA
Maximum current:  25 mA, short circuit protected
Maximum input voltage:  50 V DC
Step Functions
Source Functions  Voltage, current, resistance, frequency, temperature
Manual Step  Selectable step, change with arrow buttons
Autostep  Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
   -10 °C to +50 °C
Storage Temperature
   -20 °C to +60 °C
Dust/water resistance
   Meets IP52, IEC 529
Operating Altitude
   3000 m above mean sea level (9842 ft)

 

Safety Specifications
Agency Approvals
   CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
   136 x 245 x 63 mm (5.4“ x 9.6 x 2.5)
Weight
   1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion:  7.2V, 4400mAh, 30 Wh
Battery Life
   >8 hours typical
Battery Replacement
   Replace without opening calibrator; no tools required
Side port connections
   Pressure module connector
   USB Connector to interface to your PC
   Digital instrument (HART) connector
   Connection for optional battery charger/eliminator
Data storage capacity
   1 week of calibration procedures results
90 day specifications
   The standard specification interval for the 750 Series are 1 and 2 years.
   Typical 90 day measurement and source accuracy can be estimated by
   dividing the one year "% of reading" or "%of output" specifications by 2.
   Floor specifications, expressed as "% of full scale" or
   "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors
Degrees or % of reading
Type (α) Range °C Measure °C1 Source current Source °C Allowable current2
  1 year 2 year   1 year 2 year  
100 Ω Pt (385) -200 to 100
100 to 800
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
200 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
500 μA 0.06 °C
0.017 % + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
500 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
250 μA 0.06 °C
0.017% + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
1000 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04% + 0.10 °C
150 μA 0.06 C
0.017 % + 0.05 °C
0.12 C
0.034 % + 0.10 °C
0.1 mA to 1 mA
100 Ω Pt (3916) -200 to 100
100 to 630
0.07 °C
0.02 % +0.05 °C
0.14 °C
0.04 % +0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
100 Ω Pt (3926) -200 to 100
100 to 630
0.08 °C
0.02 % +0.06 °C
0.16 °C
0.04 % +0.12 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
10 Ω Cu (427) -100 to 260 0.2 °C 0.4 °C 3 mA 0.2 °C 0.4 °C 0.1 mA to 10 mA
120 Ω Ni (672) -80 to 260 0.1 °C 0.2 °C 1 mA 0.04 °C 0.08 °C 0.1 mA to 10 mA

1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms

Temperature, Thermocouples
Type Source °C Measure °C Source °C
    1 year 2 years 1 year 2 years
E -250 to -200 1.3 2.0 0.6 0.9
  -200 to -100 0.5 0.8 0.3 0.4
  -100 to 600 0.3 0.4 0.3 0.4
  600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
  -100 to 900 0.5 0.8 0.5 0.8
  900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1.0 0.4 0.6
  -100 to 400 0.3 0.4 0.3 0.4
  400 to 1200 0.5 0.8 0.3 0.4
  1200 to 1372 0.7 1.0 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
  -200 to 0 0.6 0.9 0.4 0.6
  0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2.0 1.0 1.5
  800 to 1000 1.0 1.5 0.8 1.2
  1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
  0 to 100 1.5 2.2 1.1 1.7
  100 to 1767 1.0 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
  0 to 200 1.5 2.1 1.1 1.7
  200 to 1400 0.9 1.4 0.9 1.4
  1400 to 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
  800 to 1200 0.8 1.2 0.7 1.0
  1200 to 1800 1.1 1.6 0.9 1.4
  1800 to 2316 2.0 3.0 1.3 2.0
L -200 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
  1000 to 2000 1.6 2.4 0.6 0.9
  2000 to 2500 2.0 3.0 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6

Description

Fluke 754 Documenting Process Calibrator - HART

The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package. 

For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.

 

Documenting Process Calibrator-HART including:

  • BC7240 battery charger
  • Li-on BP7240 battery pack
  • DPCTrack 2™ Sample Software
  • Instruction manual
  • NIST-traceable calibration report and data
  • Three sets of TP220 test probes with three sets of “extended tooth alligator clips
  • Two sets AC280 hook clips
  • C799 Soft Field Case
  • USB communication cable, Fluke 754HHC HART communications cable

Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the FlukeDPC/TRACK2™ software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.

The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.

The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART® interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator. 

  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
  • Holds up to a full week of downloaded procedures and calibration results.
  • Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
  • Easy to use
  • Three year warranty
  • Bright white dual display. Read both sourced and measured parameters simultaneously.
  • Multi lingual interface
  • Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
  • Delivered with DPC/Track Sample software.
  • Compatible with many Asset Management software packages.

HART Capabilities:

The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.

  • Requires no external box or second tool for everyday HART calibration and maintenance.
  • Offers fast HART communication.
  • Supports popular models of HART transmitters, with more devise-specific command support than any other HART field calibrator.
  • Works with multiple masters, burst mode, and multi-drop configurations.
  • Is easy to update as additional instruments are added and new HART versions are released.
  • Interrogate to determine device type, manufacturer, model, tag.
  • Reconfigure the sensor mapping of dual sensor temperature transmitters.
  • Read HART PV function and smart transmitter digital output while measuring analog mA output.
  • Read and write HART configuration functions to make field adjustments to PV range points, damping, and other top-level configuration settings.
  • Re-label smart transmitters by reading and writing to the HART tag field.

 

Manufacturer Pressure instruments Temperature instruments Coriolis instruments
ABB/Kent-Taylor 600T 658T1  
ABB/Hartmann & Braun Contrans P1,
AS 800 Series
   
Endress & Hauser CERABAR S,
CERABAR M,
DELTABAR S
TMT 1221,TMT 1821,
TMT 1621
 
Foxboro Eckardt   TI/RTT201  
Foxboro/Invensys I/A Pressure    
Fuji FCX
FCXAZ
FRC  
Honeywell ST3000 STT25T1,
STT25H1
 
Micro Motion     2000
2000 IS
9701
9712
9739
Moore Products   3441  
Rosemount 1151
2088
3001C
3051, 3051S
3044C
644
3144
3244, 3144P
 
Siemens SITRANS P DS
SITRANS P ES
   
SMAR LD301 TT3011  
Viatran I/A Pressure    
Wika UNITRANS T32H1  
Yokogawa EJA YTA 110, 310 
and 320
 

Features & Specification

More Information
Key Features Log Data, Upload data to PC, Hart communication
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Fluke Testers
Short Description
Measurement Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV        0.02%+0.005 mV  0.03%+0.005 mV
3.00000 V  0.02%+0.00005 V  0.03%+0.00005 V
30.0000 V  0.02%+0.0005 V  0.03%+0.0005 V
300.00 V  0.05%+0.05 V  0.07%+0.05 V
Voltage AC
Range 
40 to 500 Hz     
 Resolution  1 Year 2 Years 
3.000 V  0.001 V  0.5%+0.002 V  1.0%+0.004 V
30.00 V  0.01 V  0.5%+0.02 V  1.0%+0.04 V
300.0 V  0.1 V  0.5%+0.2 V  1.0%+0.2 V
Current DC
   1 Year  2 Years
30.000 mA          0.01% + 5 uA   0.015%+7 uA
110.00 mA  0.01% + 20 uA  0.015%+30 uA
Resistance
   1 Year  2 Years
10.000 Ω             0.05% + 50 mΩ  0.07%+70 mΩ
100.00 Ω  0.05% + 50 mΩ  0.07%+70 mΩ
1.0000 kΩ   0.05% + 500 mΩ  0.07%+0.5 Ω
10.000 kΩ  0.1% + 10 Ω  0.15%+15 Ω
Frequency
   Resolution  Accuracy (2 Years)
1.00 to 110.00 Hz  0.01 Hz  0.05 Hz
110.1 to 1100.0 Hz  0.1 Hz  0.5 Hz
1.101 to 11.000 kHz   0.001 kHz  0.005 kHz
11.01 to 50.00 kHz  0.01 kHz  0.05 kHz

 

Source Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV            0.01%+0.005 mV  0.015%+0.005 mV
1.00000 V  0.01%+0.00005 V  0.015%+0.0005 V
15.0000 V  0.01%+0.0005 V  0.015%+0.0005 V
Current DC
   1 Year 2 Years 
22.000 mA (Source)  0.01%+ 0.003 mA  0.02%+ 0.003 mA
Current Sink (Simulate)  0.02% + 0.007 mA  0.04% + 0.007 mA
Resistance
   1 Year  2 Years
10.000 Ω                0.01% + 10 mΩ  0.015% + 15 mΩ
100.00 Ω  0.01% + 20 mΩ  0.015% + 30 mΩ
1.0000 kΩ  0.02% + 0.2 Ω  0.03% + 0.3 Ω
10.000 kΩ  0.02% + 3 Ω  0.03% + 5 Ω
Frequency
   2 Years
0.1 to 10.99 Hz  0.01 Hz
0.01 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1.100 to 21.999 kHz  0.002 kHz
22.000 to 50.000 kHz           0.005 kHz

 

Technical Data
Data Log Functions
Measure functions:  Voltage, current, resistance, frequency, temperature, pressure
Reading rate:  1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length:  8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Rate:  4 steps/second
Trip detect:  Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage:  Selectable, 26 V
Accuracy:  10%, 18 V minimum at 22 mA
Maximum current:  25 mA, short circuit protected
Maximum input voltage:  50 V DC
Step Functions
Source Functions  Voltage, current, resistance, frequency, temperature
Manual Step  Selectable step, change with arrow buttons
Autostep  Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
   -10 °C to +50 °C
Storage Temperature
   -20 °C to +60 °C
Dust/water resistance
   Meets IP52, IEC 529
Operating Altitude
   3000 m above mean sea level (9842 ft)

 

Safety Specifications
Agency Approvals
   CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
   136 x 245 x 63 mm (5.4“ x 9.6 x 2.5)
Weight
   1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion:  7.2V, 4400mAh, 30 Wh
Battery Life
   >8 hours typical
Battery Replacement
   Replace without opening calibrator; no tools required
Side port connections
   Pressure module connector
   USB Connector to interface to your PC
   Digital instrument (HART) connector
   Connection for optional battery charger/eliminator
Data storage capacity
   1 week of calibration procedures results
90 day specifications
   The standard specification interval for the 750 Series are 1 and 2 years.
   Typical 90 day measurement and source accuracy can be estimated by
   dividing the one year "% of reading" or "%of output" specifications by 2.
   Floor specifications, expressed as "% of full scale" or
   "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors
Degrees or % of reading
Type (α) Range °C Measure °C1 Source current Source °C Allowable current2
  1 year 2 year   1 year 2 year  
100 Ω Pt (385) -200 to 100
100 to 800
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
200 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
500 μA 0.06 °C
0.017 % + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
500 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
250 μA 0.06 °C
0.017% + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
1000 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04% + 0.10 °C
150 μA 0.06 C
0.017 % + 0.05 °C
0.12 C
0.034 % + 0.10 °C
0.1 mA to 1 mA
100 Ω Pt (3916) -200 to 100
100 to 630
0.07 °C
0.02 % +0.05 °C
0.14 °C
0.04 % +0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
100 Ω Pt (3926) -200 to 100
100 to 630
0.08 °C
0.02 % +0.06 °C
0.16 °C
0.04 % +0.12 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
10 Ω Cu (427) -100 to 260 0.2 °C 0.4 °C 3 mA 0.2 °C 0.4 °C 0.1 mA to 10 mA
120 Ω Ni (672) -80 to 260 0.1 °C 0.2 °C 1 mA 0.04 °C 0.08 °C 0.1 mA to 10 mA

1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms

Temperature, Thermocouples
Type Source °C Measure °C Source °C
    1 year 2 years 1 year 2 years
E -250 to -200 1.3 2.0 0.6 0.9
  -200 to -100 0.5 0.8 0.3 0.4
  -100 to 600 0.3 0.4 0.3 0.4
  600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
  -100 to 900 0.5 0.8 0.5 0.8
  900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1.0 0.4 0.6
  -100 to 400 0.3 0.4 0.3 0.4
  400 to 1200 0.5 0.8 0.3 0.4
  1200 to 1372 0.7 1.0 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
  -200 to 0 0.6 0.9 0.4 0.6
  0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2.0 1.0 1.5
  800 to 1000 1.0 1.5 0.8 1.2
  1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
  0 to 100 1.5 2.2 1.1 1.7
  100 to 1767 1.0 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
  0 to 200 1.5 2.1 1.1 1.7
  200 to 1400 0.9 1.4 0.9 1.4
  1400 to 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
  800 to 1200 0.8 1.2 0.7 1.0
  1200 to 1800 1.1 1.6 0.9 1.4
  1800 to 2316 2.0 3.0 1.3 2.0
L -200 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
  1000 to 2000 1.6 2.4 0.6 0.9
  2000 to 2500 2.0 3.0 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA