Time Electronics 1017 Voltage, Current, Resistance Calibrator

£1,860.00

Voltage
Ranges & Accuracy

0 to 9.99999mV in 10nV steps / ± 0.02% of setting ± 0.005% of range
0 to 99.9999mV in 100nV steps / ± 0.01% of setting ± 0.004% of range
0 to 999.999mV in 1μV steps / ± 0.005% of setting ± 0.002% of range
0 to 9.99999V in 10μV steps / ± 0.005% of setting ± 0.002% of range
0 to 99.9999V in 100μV steps / ± 0.01% of setting ± 0.004% of range

Output Resistance

10mV and 100mV: 10Ω. 1V and 10V: <150mΩ. 100V: <1Ω.

Drive Current Max

10 and 100mV: as 10Ω output resistance. 1V and 10V: 150mA; 100V: 10mA.

Current
Range & Accuracy


0 to 99.9999mA in 0.1μA steps / ± 0.02% of setting ± 0.004% of range

Drive Voltage Max

10V

Resistance
Range & Accuracy


0 to 9.99999kΩ in 0.01Ω steps / ± 0.05% of setting ± 0.003% of range

Power Rating

0.25W per resistor

Residual resistance

Less than 200mΩ

   

Deviation Control (V&I)

0% to 0.999% in 0.001% steps. Deviation accuracy: V & I output, 0.5%.

Temperature Coefficient

<5ppm per °C

Long Term Stability <5ppm/day, <15ppm/90day, <25ppm/year

Short Term Stability – Noise

10mV range: <0.2μV/sec, <0.3μV/10sec, <0.4μV/min
100mV range: <0.2μV/sec, <0.4μV/10sec, <0.6μV/min
1V range: <0.2μV/sec, <0.5μV/10sec, <1.5μV/min
10V range: <1.0μV/sec, <2.0μV/10sec, < 8.0μV/min
100V range: <40μV/sec, <100μV/10sec, <500μV/min
100mA range: <0.2μA/sec, <0.4μA/10sec, <1.0μA/min

Warm-up/Settling Time

Warm-up: <10 mins to full accuracy. Settling: < 0.5 secs, 100V range 5 secs.

Output Connections The output is via low thermal emf terminals (0.2μV/°C). A mains earth terminal is provided for screening purposes. Output polarity can be selected by a switch on the front panel.
Power Supply The 1017 can be powered continuously from a 230V 50/60 Hz (110V to order) mains supply, or from the internal rechargeable NiCad battery pack.
Battery Level Indicator A front panel display provides a continuous indication of the battery state.

The above accuracies are independent of thermal emfs which can be 2μV or more depending on the type of leads and connections used.

 

Operating Temperature 0 to 50°C (32 to 120°F). 15 to 25°C for optimum performance.
Operating Humidity 10 to 90% non-condensing 25°C (77°F)

Dimensions

W250x H119 x D314mm

Weight

2.4kg

Optional Extras

Calibration Certificates – traceable to NPL and UKAS

Country of Origin UK

Description

A high performance portable multifunction calibrator with voltage, current, and resistance ranges. The 1017 combines precision with simple operation, making it suitable for use in the workshop or field. Constructed in a compact and durable plastic case with a tilt stand/carry handle it takes up minimal bench space and is easily transportable.

A portable multifunction calibrator with voltage, current, and resistance ranges. This model is suitable for calibrating a wide range of instruments including thermocouples, transducers, transmitters, and platinum resistance thermometers.

• DC Voltage 10nV to 100V
• DC Current 100nA to 100mA
• Resistance 10mΩ to 10kΩ
• Accuracy 0.005% of setting + 0.002% of rng
• 1ppm setting resolution
• Noise <2ppm (0.1 to 1Hz)
• Stability <5ppm/day, <25ppm/yr
• Residual resistance <200mΩ
• Deviation control – Voltage and Current
• Battery or mains operation
• 12 hours typical use between charges
• W250x H119 x D314mm, weight 2.4kg

Five DC voltage ranges from 10mV to 100V full scale are available, each with a 6-digit (1ppm) resolution. The DC current range is 100mA full scale with a 100nA (1ppm) resolution. Resistance from 0.01Ω to 10kΩ is available 0.01Ω steps.

The 1017 can be powered from mains supply or by the internal rechargeable battery pack. Battery operation enables good performance where earth loop and noise pick-up occurs. When the calibrator is plugged into the mains supply the internal batteries will automatically start to recharge. If unplugged from the mains during operation the internal batteries will continue to power the instrument. Full charge allows 12 hours typical use. The battery condition monitored by a meter on the front panel. 

Stability vs Temperature and Time
Outstanding performance is achieved by the use of special computer selected reference diodes and the latest in resistor technology. The special low-thermal emf terminals reduce errors when working with microvolt signals.

Digital Deviation Control
Allows the output to be increased/decreased in % terms from 0 to +/-0.999%. This provides a direct read-out of error and simplifies the recording results for calibration certificates. It enables the user to immediately see if the unit under test is within specification.

Applications
The 1017 is suitable for calibrating and simulating a wide range of instruments including thermocouples, transducers, 4-20mA and 0-10V transmitters, and platinum resistance thermometers.

As part of the Time Electronics multifunction calibrators range the 1017 is designed for accuracy and reliability, whilst maintaining ease of use.
It is a versatile calibration instrument that provides a high performance solution to voltage, current and resistance applications

Features & Specification

More Information
Key Features DC Volts 10nV to 100V, 0.005% Accuracy, Resistance 10mΩ to 10KΩ
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Time Electronics
Short Description

Voltage
Ranges & Accuracy

0 to 9.99999mV in 10nV steps / ± 0.02% of setting ± 0.005% of range
0 to 99.9999mV in 100nV steps / ± 0.01% of setting ± 0.004% of range
0 to 999.999mV in 1μV steps / ± 0.005% of setting ± 0.002% of range
0 to 9.99999V in 10μV steps / ± 0.005% of setting ± 0.002% of range
0 to 99.9999V in 100μV steps / ± 0.01% of setting ± 0.004% of range

Output Resistance

10mV and 100mV: 10Ω. 1V and 10V: <150mΩ. 100V: <1Ω.

Drive Current Max

10 and 100mV: as 10Ω output resistance. 1V and 10V: 150mA; 100V: 10mA.

Current
Range & Accuracy


0 to 99.9999mA in 0.1μA steps / ± 0.02% of setting ± 0.004% of range

Drive Voltage Max

10V

Resistance
Range & Accuracy


0 to 9.99999kΩ in 0.01Ω steps / ± 0.05% of setting ± 0.003% of range

Power Rating

0.25W per resistor

Residual resistance

Less than 200mΩ

   

Deviation Control (V&I)

0% to 0.999% in 0.001% steps. Deviation accuracy: V & I output, 0.5%.

Temperature Coefficient

<5ppm per °C

Long Term Stability <5ppm/day, <15ppm/90day, <25ppm/year

Short Term Stability – Noise

10mV range: <0.2μV/sec, <0.3μV/10sec, <0.4μV/min
100mV range: <0.2μV/sec, <0.4μV/10sec, <0.6μV/min
1V range: <0.2μV/sec, <0.5μV/10sec, <1.5μV/min
10V range: <1.0μV/sec, <2.0μV/10sec, < 8.0μV/min
100V range: <40μV/sec, <100μV/10sec, <500μV/min
100mA range: <0.2μA/sec, <0.4μA/10sec, <1.0μA/min

Warm-up/Settling Time

Warm-up: <10 mins to full accuracy. Settling: < 0.5 secs, 100V range 5 secs.

Output Connections The output is via low thermal emf terminals (0.2μV/°C). A mains earth terminal is provided for screening purposes. Output polarity can be selected by a switch on the front panel.
Power Supply The 1017 can be powered continuously from a 230V 50/60 Hz (110V to order) mains supply, or from the internal rechargeable NiCad battery pack.
Battery Level Indicator A front panel display provides a continuous indication of the battery state.

The above accuracies are independent of thermal emfs which can be 2μV or more depending on the type of leads and connections used.

 

Operating Temperature 0 to 50°C (32 to 120°F). 15 to 25°C for optimum performance.
Operating Humidity 10 to 90% non-condensing 25°C (77°F)

Dimensions

W250x H119 x D314mm

Weight

2.4kg

Optional Extras

Calibration Certificates – traceable to NPL and UKAS

Country of Origin UK

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA