Fluke a3001 FC iFlex Current Clamp Kit

£586.80
Detailed Specifications
AC Voltage
Range 1 Resolution Accuracy 2 3 4
45 Hz to 500 Hz 500 Hz to 1 kHz
600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3
6.000 V 0.001 V
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V
1 All AC voltage ranges are specified from 1 % of range to 100 % of range. 
2 Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor up to 3. 
Do not exceed 107 V-Hz.

 

DC Voltage, Continuity, Resistance, Diode Test and Capacitance
Function Range Resolution Accuracy
mV 600.0 mV 0.1 mV 0.09 % + 2
V 6.000 V 0.001 V 0.09 % + 2
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V 0.15 % + 2
Ω 600 Ω 1 Ω Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 μs or longer.
Ω 600.0 Ω 0.1 Ω 0.5 % + 2
6.000 kΩ 0.001 kΩ 0.5 % + 1
60.00 kΩ 0.01 kΩ
60.00 kΩ 0.1 kΩ
600.0 kΩ 0.001 MΩ
50.00 MΩ 0.01 MΩ 1.5 % + 3
Diode test 2.000 V 0.001 V 1 % + 2
µF 1000 nF 1 nF 1.2 % + 2
10.00 μF 0.01 μF
100.0 μF 0.1 μF
9999 μF1 1 μF 10 % typical
1 In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.

 

AC and DC Current
Function Range 1 Resolution Accuracy
mA AC 
(45 Hz to 1 kHz)
60.00 mA 0.01 mA 1.5 % + 3
400.0 mA 3 0.1 mA
mA DC 2 60.00 mA 0.01 mA 0.5 % + 3
400.0 mA 3 0.1 mA
1 All AC current ranges are specified from 5 % of range to 100 % of range. 
2 Input burden voltage (typical): 400 mA input 2 mV/mA. 
3 400.0 mA accuracy specified up to 600 mA overload.

 

Frequency
Range Resolution Accuracy 1
99.99 Hz 0.01 Hz 0.1 % + 1
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
1 Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.

 

Input Characteristics
Function Overload 
protection
Input impedance 
(nominal)
Common mode rejection ratio (1 kΩ unbalance) Normal mode rejection
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
1100 V RMS > 10 MΩ < 100 pF > 60 dB, DC to 60 Hz  
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
Open circuit test voltage Full scale voltage Typical short circuit current
To 6 MΩ 50 MΩ
1100 V RMS < 2.7 V DC < 0.7 V DC < 0.9 V DC < 350 mA
1100 V RMS < 2.7 V DC 2.000 V DC < 1.1 mA

 

mA Function
Overload Protection Fused, 44/100 A, 1000 V FAST Fuse
Overload 600 mA overload for 2 minutes maximum, 10 minutes rest minimum

 

MIN/MAX Recording Accuracy
DC functions The specified accuracy of the measurement function ± 12 counts for changes > 350 mS in duration.
AC functions The specified accuracy of the measurement function ± 40 counts for changes > 900 mS in duration.

 

General Specifications
Maximum voltage between any terminal and earth ground 1000 V DC or AC RMS
Ω fuse protection from A inputs 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
Display (LCD) Update rate: 4/sec 
Volts, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts
Battery type Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life 250 hours minimum
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20m
Obstructed, sheetrock wall: Up to 6.5m
Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m
Temperature Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Relative humidity 0 % to 90 % (0 °C to 35 °C)
0 % to 75 % (35 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C)
Altitude Operating: 2,000 m
Storage: 12,000 m
Electromagnetic compatibility
EMI, RFI, EMC, RF
EN 61326-1:2006, EN 61326-2-2:2006
ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
Safety compliance ANSI/ISA 61010-1 / (82.02.01): 3rd edition 
CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL 61010-1: 3rd edition 
IEC/EN 61010-1:2010
Certifications CSA, FCC, CE
Ingress Protection (IP) rating IP54
Pollution degree 2
Size (H x W x L) 4.75 cm x 9.3 cm x 20.7 cm
(1.87 in x 3.68 in x 8.14 in)
Weight 340 g (12 oz)
Warranty Three years

Note: Not compatible with Fluke CNX test tools

Fluke a3001 FC Wireless iFlex AC Current Module

 

Specifications
Range 0.5 A to 2500 A AC
Resolution 0.1 A
Accuracy 3 % ± 5 digits (5 Hz to 500 Hz)
Crest factor (50 Hz/60 Hz) 3.0 at 1100 A, 2.5 at 1400 A, 1.42 at 2500 A, add 2 % for C.F. > 2
LCD w/backlight 3½ digits
Log rate/interval 1 sec minimum/adjustable by PC or front panel
Battery type 2 AA, NEDA 15 A, IEC LR6
Battery life 400 hours
Memory Record up to 65,000 readings
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20 m
Obstructed, sheetrock wall: Up to 6.5 m
Obstructed, concrete wall or steel electrical enclosure: Up to 3.5 m
Operating temperature -10 °C to +50 °C
Storage temperature -40 °C to +60 °C
Temperature coefficient Add 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
Operating humidity 90 % at 35 °C, 75 % at 40 °C, 45 % at 50 °C
Altitude Operating: 2,000 m
Storage: 12,000 m
EMC EN 61326-1:2006
Safety compliance IEC 61010-1, 600 V CAT IV/1000 V CAT III, 3rd edition
Safety rating CAT IV 600 V, CAT III 1000 V
Certifications CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
Ingress Protection (IP) rating IP42
Pollution degree 2
Jaw opening 25.4 cm (10 in) coil
Size (HxWxD) 16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in)
Weight .22 kg (8 oz)
Warranty Three years

Note: Not compatible with Fluke CNX test tools

Description

Fluke a3001 FC Wireless iFlex AC Current Clamp Kit

Part of the Fluke Connect™ team of wireless test tools

 

Connect the a3001 FC Wireless AC Current Module and flexible current probe to your test point, and view the results up to 20 meters away on the 3000 FC Wireless Digital Multimeter.

 

Fluke FC Wireless a3001 Comes Supplied With:

  • Fluke 3000 FC Series Wireless Multimeter
  • Fluke a3001 FC Wireless iFlex AC Current Module
  • TL175 Test Leads
  • AC175 Alligator Clips
  • iFlex i2500-10 Flexible Current Probe
 

You’ll save time, with less run-around collecting multiple measurements. Use multiple modules for measurements on three-phase systems, as a stand-alone measurement tool or combine with other Fluke Connect™ modules as a system for multiple measurements.

From short distances, you can even view readings from modules through closed electrical panels. Plus no more writing down data as the FC Wireless AC Current Module captures up to 65,000 sets of time stamped min/max/avg readings, using the optional PC adapter. Fluke Connect wireless test tools also offer increased safety by letting you view readings in a separate location from the test point.

Now you can take readings on moving machinery, with only the measurement module in harm’s way.

 

The Fluke a3001 FC Wireless iFlex® AC Current Clamp Kit provides measurement of many common electrical parameters

The Fluke 3000 FC Series Wireless Multimeter with the Fluke Connect™ app has all the essentials for convenient test and measurement troubleshooting:

  • AC and DC voltage measurements to 1000V
  • AC and DC current with 0.01 mA resolution
  • Continuity, resistance, diode test, capacitance and frequency measurements
  • MIN/MAX recording
  • CAT III 1000 V, Cat IV 600 V; IP54
 

The a3001 Wireless iFlex® Current Module allows you to measure around awkward sized conductors or get into tight spaces and makes wire access easier:

  • Measure up to 2500 A AC
  • Use as a standalone meter or as part of the system
  • Logging function for recording and saving up to 65,000 readings
  • Inrush function

Features & Specification

More Information
Key Features iFlex wireless kit, Test leads included, Cutting edge technology, Tough, durable material
Product Stock Status Next Day Delivery (subject to availability)
Manufacturer Fluke Testers
Short Description
Detailed Specifications
AC Voltage
Range 1 Resolution Accuracy 2 3 4
45 Hz to 500 Hz 500 Hz to 1 kHz
600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3
6.000 V 0.001 V
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V
1 All AC voltage ranges are specified from 1 % of range to 100 % of range. 
2 Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor up to 3. 
Do not exceed 107 V-Hz.

 

DC Voltage, Continuity, Resistance, Diode Test and Capacitance
Function Range Resolution Accuracy
mV 600.0 mV 0.1 mV 0.09 % + 2
V 6.000 V 0.001 V 0.09 % + 2
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V 0.15 % + 2
Ω 600 Ω 1 Ω Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 μs or longer.
Ω 600.0 Ω 0.1 Ω 0.5 % + 2
6.000 kΩ 0.001 kΩ 0.5 % + 1
60.00 kΩ 0.01 kΩ
60.00 kΩ 0.1 kΩ
600.0 kΩ 0.001 MΩ
50.00 MΩ 0.01 MΩ 1.5 % + 3
Diode test 2.000 V 0.001 V 1 % + 2
µF 1000 nF 1 nF 1.2 % + 2
10.00 μF 0.01 μF
100.0 μF 0.1 μF
9999 μF1 1 μF 10 % typical
1 In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.

 

AC and DC Current
Function Range 1 Resolution Accuracy
mA AC 
(45 Hz to 1 kHz)
60.00 mA 0.01 mA 1.5 % + 3
400.0 mA 3 0.1 mA
mA DC 2 60.00 mA 0.01 mA 0.5 % + 3
400.0 mA 3 0.1 mA
1 All AC current ranges are specified from 5 % of range to 100 % of range. 
2 Input burden voltage (typical): 400 mA input 2 mV/mA. 
3 400.0 mA accuracy specified up to 600 mA overload.

 

Frequency
Range Resolution Accuracy 1
99.99 Hz 0.01 Hz 0.1 % + 1
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
1 Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.

 

Input Characteristics
Function Overload 
protection
Input impedance 
(nominal)
Common mode rejection ratio (1 kΩ unbalance) Normal mode rejection
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
1100 V RMS > 10 MΩ < 100 pF > 60 dB, DC to 60 Hz  
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
Open circuit test voltage Full scale voltage Typical short circuit current
To 6 MΩ 50 MΩ
1100 V RMS < 2.7 V DC < 0.7 V DC < 0.9 V DC < 350 mA
1100 V RMS < 2.7 V DC 2.000 V DC < 1.1 mA

 

mA Function
Overload Protection Fused, 44/100 A, 1000 V FAST Fuse
Overload 600 mA overload for 2 minutes maximum, 10 minutes rest minimum

 

MIN/MAX Recording Accuracy
DC functions The specified accuracy of the measurement function ± 12 counts for changes > 350 mS in duration.
AC functions The specified accuracy of the measurement function ± 40 counts for changes > 900 mS in duration.

 

General Specifications
Maximum voltage between any terminal and earth ground 1000 V DC or AC RMS
Ω fuse protection from A inputs 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
Display (LCD) Update rate: 4/sec 
Volts, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts
Battery type Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life 250 hours minimum
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20m
Obstructed, sheetrock wall: Up to 6.5m
Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m
Temperature Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Relative humidity 0 % to 90 % (0 °C to 35 °C)
0 % to 75 % (35 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C)
Altitude Operating: 2,000 m
Storage: 12,000 m
Electromagnetic compatibility
EMI, RFI, EMC, RF
EN 61326-1:2006, EN 61326-2-2:2006
ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
Safety compliance ANSI/ISA 61010-1 / (82.02.01): 3rd edition 
CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL 61010-1: 3rd edition 
IEC/EN 61010-1:2010
Certifications CSA, FCC, CE
Ingress Protection (IP) rating IP54
Pollution degree 2
Size (H x W x L) 4.75 cm x 9.3 cm x 20.7 cm
(1.87 in x 3.68 in x 8.14 in)
Weight 340 g (12 oz)
Warranty Three years

Note: Not compatible with Fluke CNX test tools

Fluke a3001 FC Wireless iFlex AC Current Module

 

Specifications
Range 0.5 A to 2500 A AC
Resolution 0.1 A
Accuracy 3 % ± 5 digits (5 Hz to 500 Hz)
Crest factor (50 Hz/60 Hz) 3.0 at 1100 A, 2.5 at 1400 A, 1.42 at 2500 A, add 2 % for C.F. > 2
LCD w/backlight 3½ digits
Log rate/interval 1 sec minimum/adjustable by PC or front panel
Battery type 2 AA, NEDA 15 A, IEC LR6
Battery life 400 hours
Memory Record up to 65,000 readings
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20 m
Obstructed, sheetrock wall: Up to 6.5 m
Obstructed, concrete wall or steel electrical enclosure: Up to 3.5 m
Operating temperature -10 °C to +50 °C
Storage temperature -40 °C to +60 °C
Temperature coefficient Add 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
Operating humidity 90 % at 35 °C, 75 % at 40 °C, 45 % at 50 °C
Altitude Operating: 2,000 m
Storage: 12,000 m
EMC EN 61326-1:2006
Safety compliance IEC 61010-1, 600 V CAT IV/1000 V CAT III, 3rd edition
Safety rating CAT IV 600 V, CAT III 1000 V
Certifications CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
Ingress Protection (IP) rating IP42
Pollution degree 2
Jaw opening 25.4 cm (10 in) coil
Size (HxWxD) 16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in)
Weight .22 kg (8 oz)
Warranty Three years

Note: Not compatible with Fluke CNX test tools

Delivery

Updated January 2024

Please note, our cut-off time is 2:00pm

Orders requiring calibration 

Whilst we take the greatest care to get all orders dispatched and delivered to you in the stated time frame, orders received PM and requiring calibration may on occasion miss dispatch that day, according to the complexity of the calibration.
Be assured they will ship on the next dispatch this in most cases should not add more than 24 hours to your delivery.

Bank/Public Holidays

Predicted delivery times are based on regular trading days Monday to Friday. Public holidays, bank holidays and weekends are outside the usual scope
of our dispatch and subject to activities/holidays of our carriers and therefore may incur additional days lead time to your order.

*UK Mainland (Revised March 2021)

Service Price
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value OVER £75.00 ex VAT FREE
Standard FREE 2-3 Day Courier Delivery – UK Mainland only - Zones A & B - Order value UNDER £75.00 ex VAT £4.95
Next WORKING Day (excludes Saturdays, Sundays and public holidays)  – UK Mainland only - Zones A & B  - ANY order value - Order by 2:00pm £7.75
Saturday ANYTIME Delivery - Order by 2:00pm £9.95
 

 

 

 

 

Delivery Surcharges

Because of the high cost of carriage, we regretfully need to apply a surcharge on all deliveries made in the following regions:

  • Zones A & B  - Mainland delivery rates apply. 
  • Zone C - Scottish Highlands (Charges are calculated at the checkout).
  • Zone D - Outer Hebrides, Orkney Isles, Shetland Isles, Isle of Man, Scilly Isles, and Channel Islands (Charges are calculated at the checkout).
  • Zone E - Northern Ireland (Charges are calculated at the checkout).
  • Zone F - Republic of Ireland (Charges are calculated at the checkout).

Offshore Services

All offshore deliveries are a 3-4 day service as standard. This includes Channel Islands, Northern & Southern Ireland, Scottish Islands, Isle of Man and the Scilly Isles.

International delivery

International delivery charges are automatically added to your order at the checkout. For any questions regarding international deliveries please contact us at sales@isswww.co.uk

Export 

All export inquiries should be directed to the export division before placement of order.

If no shipping option is available to your selected country please call us FREE on 0800 0224 060 to discuss your requirements.
All stated prices are subject to VAT at the current rate where applicable.
E&OA